Section 1 Product overview

The Intellical PHC101 series pH probes are digital, combination electrodes that measure the pH of
wastewater, drinking water and general water samples. The probes have a non-refillable, gel-filled
electrolyte and a built-in temperature sensor. The open reference junction gives an optimum
electrical connection between the sample and the electrolyte and does not become clogged. The
standard probes are for laboratory use. The rugged probes are for field use. Refer to
Note: Do not use the probe to measure the pH of organic solvents or samples with a pH less than 2.
Figure 1 Probe overview
1 Shroud (rugged model)
2 Probe storage cap
3 Glass bulb and temperature sensor
4 Reference junction
5 Protective tape

Section 2 Specifications

Specifications are subject to change without notice.
Specifications
Probe type
pH range
pH accuracy
Reference type
6 Locking ring (rugged model)
7 Rugged probe
8 Probe soaker bottle holder or storage cap
9 Standard probe
10 Probe soaker bottle with storage solution
Details
Digital combination pH probe with non-refillable gel electrolyte
and built-in temperature sensor
2 to 14 pH
±0.02 pH
Ag/AgCl