Omron HV-F013 Instruction Manual

Omron HV-F013 Instruction Manual

Electronic nerve stimulator
Hide thumbs Also See for HV-F013:

Advertisement

Quick Links

Thank you for purchasing the OMRON Electronic Nerve
Stimulator.
OMRON Electronic Nerve Stimulator is intended for the relief
In order to use the unit safely, read the complete manual carefully
of pain associated with sore or aching muscles of the lower back,
before using the unit for the fi r st time.
arms, legs, feet and shoulder due to strain from exercise or normal
Keep this instruction manual in a convenient place or store with
the unit for future reference.
household and work activities. It should be applied to normal,
This box contains the following components:
healthy, dry, clean skin of adult patients.
• 1 Main Unit
• 1 Electrode Cord
This unit is intended to be operated by adults who can understand
• 2 Long Life Pads
this instruction manual. It is not for professional use in hospitals
• 1 Pad Holder
or other medical facilities, it is intended for home use only.
• 2
Batteries (LR03)
• 1 Instruction Manual
• 1 Quick Start Guide / Pad Placement Guide
it
Scientifi c theory suggests that electrical stimulation therapy may
It is designed to reduce and
work in several ways:
relieve muscle and joint pain, stiffness and numbness in the
back, arms, legs, shoulder and foot by applying electrical nerve
1. The gentle electrical pulses move through the skin to nearby
stimulation to the surface of the skin near the site of the pain.
nerves to block or shut out the pain message from ever
reaching the brain from the source of the pain.
2. The gentle electrical pulses increase the production of the
It can be successfully used in conjunction with any other
body's natural pain killer, such as endorphins.
pain treatment or medication.
3. Furthermore, it is thought that the electrical stimulation
This type of therapy, called Transcutaneous Electrical Nerve
improves blood circulation as well. Muscles contract and
Stimulator (TENS), has been used for over 30 years by medical
relax with the flow of the electrical stimulation. With repeated
professionals such as physical therapists and chiropractors.
contracting and relaxing, the blood fl o ws in and out and the
blood circulation is improved.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Improper use may cause danger resulting in death or serious injury. These are situations in which the device should not be
used.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient
or damage to the equipment or other property.
Do not use this unit with these other devices:
If you have a
implanted defibrillator, or other
or electronic device. Such use could cause electric shock,
bums, electrical interference or death.
(2)
this device
device.
(3) Together with a
such as an artificial heart or lung or respirator.
(4) For Hospitals and Clinics, in the presence of or
which may not operate properly when the electrical stimulation device is in use.
(5) For Hospitals and Clinics, Simultaneous
of a patient
stimulator electrodes and possible damage to the stimulator.
(6) For Hospitals and Clinics, operation
(e.g. Jm) to
the stimulator output.
The unit may cause lethal rhythm disturbances in certain susceptible individuals. If you have had a recent surgical procedure, the
stimulation may disrupt the healing process.
Do not use on
because this device has not been evaluated for pediatric use.
Keep out of the reach of
because the electrode cord could cause strangulation.
Persons incapable of expressing their thoughts or intentions.
Persons incapable of operating the unit by themselves.
Use caution if you have a tendency to bleed internally, such as following an injury or fracture.
If you have suspected or diagnosed heart disease, you should follow precautions recommended by your physician.
If
you should follow precautions recommended by your physician.
Use caution if stimulation is applied
This device is not intended for use by persons (including children) whose physical, sensory or mental abilities are different or
reduced, or lack of experience or knowledge unless such persons are afforded supervision or training to operate the appliance by a
person responsible for their safety. Children should be supervised to ensure they do not use the equipment as a toy.
The
the
or any area of the
Both sides of the
front and back), or
introduction of electrical current may cause rhythm
The
or any area of the
because this could
disturbances which could be lethal.
cause severe muscle spasms resulting in closure of the
airway, difficulty in breathing, or adverse effects on
heart rhythm or blood pressure.
On the
or on
this may cause cardiac disturbance.
On the
this may cause cardiac disturbance.
or
or
thrombophlebitis and thrombosis), or on top of or close to
or over areas of
DO NOT USE THIS UNIT DURING THESE ACTIVITIES
When in the bath or shower;
While sleeping;
While driving, operating machinery, or during any activity in which electrical stimulation can put you at risk of injury.
omRon
INSTRUCTION MANUAL
Electronic Nerve Stimulator
HV-F013
All for Healthcare
INTRODUCTION
Intended Use
Medical purpose
Intended User
"AAN'
HOW Electronic Nerve Stimulator WORKS
What is it?
How does
The OMRON Electronic Nerve Stimulator unit is a portable
device designed for adults only.
Any of the modes can safely be used on body parts or pains
described in this manual or Quick Start Guide/Pad Placement
Guide. Just find one that feels good and is comfortable on your
pain.
IMPORTANT SAFETY PRECAUTIONS AND WARNINGS
� It is important that you read all the warnings and precautions included in this instruction manual because they are intended to
� keep you safe, prevent injury and avoid a situation that could result in damage to the unit.
A Danger
A Warning
&.Caution
A Danger
(!)
pacemaker,
implanted metallic
Do not use
while using another TENS
life-supporting medical electronic device
when attached to the body, electronic monitoring equipment
connection
to a high frequency surgical ME equipment
in close proximity
a shortwave or microwave therapy ME equipment
DO NOT USE THIS UNIT UNDER THESE CONDITIONS
A Consult with your physician before using this unit.
DO NOT USE ON THESE INDIVIDUALS
A Pregnant women.
A
children or infants
A
young children
A
A
A
A
A
you have suspected or diagnosed epilepsy,
A
over the menstruating uterus.
A
NEVER APPLY THE PADS TO THESE BODY AREAS:
A
®
®
head,
mouth,
face.
.
neck
throat
@
®
®
Near the heart,
genital area.
®
Open wounds
rashes
over swollen, red, infected or inflamed areas or skin eruptions
cancerous lesions,
A
A
A
If you have had medical or physical treatment for your
If your pain does not improve, becomes seriously chronic or severe, or continues for
The mere existence of pain functions as a very important warning telling us that something is wrong. Therefore, if you suffer from
any serious illness,
unit.
If you have any infectious disease or illness,
Do not plug this cord into any other device that is not an OMRON HV-F013.
of this unit is allowed.
the unit.
Apply pads to normal, healthy, dry, clean skin (of adult patients) because it may otherwise disrupt the healing
process.
If you experience any skin irritation or redness after a session, do not continue stimulation in that area of the skin.
LL Therapy won't work with just one pad.
&.
not be effective.
LL
LL
LL
infection.
To avoid damage to the adhesive surface of the pads,
L'.h
1 - Pad Placement] [GET STARTED WITH YOUR THERAPY -STEP 1 - Pad Placement (continued)]).
self-adhesive pads will adhere to your skin.
LL If
immediately stop using the unit.
intended for.
bathroom. This will damage the unit.
10
C - 40
C, 30% - 80% relative humidity.
Do not use the unit without proper lighting. You may not be
able to operate unit successfully.
Do not wash the electrode cord.
the jack on the unit.
LL
LL
LL When pulling out the cord from the unit,
LL
LL The long-term effects of electrical stimulation are unknown.
Apply stimulation to only normal, intact, clean, dry, and healthy skin.
LL TENS is not effective in treating the original source or cause of the pain, including headache.
TENS is not a substitute for pain medications and other pain management therapies.
LL TENS devices do not cure disease or injuries.
LL TENS is a symptomatic treatment and, as such, suppresses the sensation of pain that would otherwise serve as a protective
mechanism.
LL Effectiveness is highly dependent upon patient selection by a practitioner qualifi e d in the management of pain patients.
(e.g. cardiac monitors, ECG alarms),
eyes and to your head and face.
may result in bums at the site of the
that region may become exhausted and sore.
may produce instability to
Main Unit
Front
Electrode Cord
1.
2. Ten intensity levels (1 low to 10 high).
3.
(durable, re-usable, washable, up to 150 uses). (Only use OMRON manufactured pads or cord with this unit).
4. Automatic
let you clearly control your therapy with mode, intensity level.
simultaneously (lateral or
because the
6.
lights up to remind you to replace the batteries.
7.
blinks if pad(s) falls off.
at the same time because
Each light corresponds to a mode.
(@[GET STARTED WITH YOUR THE � Y - � TEP 2
SELECT I OF
MODES] and the Back of Mam Umt)
at the same time because
Select mode and mtens1ty accordmg to your needs.
(such as varicose veins, phlebitis,
1.
Press Jtt./'ff to choose mode.
(@[GET STARTED WITH YOUR THE RA PY -STEP 2 -
SELECT I OF 5 MODES])
2. After selectmg the mode:
Press Jt,,. for higher intensity.
Press 'ff for lower intensity.
(@[GET STARTED WITH YOUR THERAPY -STEP 3 -
SELECT INTENSITY LEVEL (I LOW - 10 HIGH)])
°
°
PAIN MANAGEMENT WARNINGS
A
A
consult with your physician.
A
A
DO NOT ALTER THE UNIT
A
A No modification
A Use this unit only with the pads, cable, and accessories recommended
Treatment
Intensity
5
10
Programs
WARNINGS REGARDING THE PADS
Levels
A
A
6602358-BE
Issue Date: 2021-06-01
PRECAUTIONS REGARDING THE PADS
Lt. Do not move the pads to another location while the unit is on.
Make sure the components
Lt. Pad should not touch any metal object,
Do not overlap pads
Gel pads may also stick together
Do not apply pads with wet hands.
Do not leave pads attached to the skin after treatment.
Lt. The pads are for the exclusive use of a single patient. Do not share pads with another person.
Lt. Do not bend or fold
Lt.
Always place clean pads in accordance with illustrations
Lt. Do not apply ointment or any solvent
Lt. Place pads at least 2.5 cm apart for optimal results.
work?
CAUTION WHILE USING UNIT
MAIN UNIT
the unit is not functioning properly
Lt. Do not use for any other purpose
Lt. Do not place in a room with high humidity,
using
Lt.
CORD
A
Lt. Do not insert the electrode plug into any place other
Do not pull on the electrode cord
Do not bend or pull the end of the cord.
pull.
Replace the cord when broken or damaged.
General Precautions
Lt.
Lt.
Lt. You may experience skin irritation or hypersensitivity
Lt. Keep unit away from young children.
Possible Adverse Reactions
A You should stop using the unit and consult with your physician if
A You may experience skin irritation and burns
A You may experience headache and other painful sensations
Lt. Do not use to treat one region for extended periods of time (more than 30 minutes a session, up to 3 times/day)
PACKAGE CONTENTS
omoon
0@
FEATURES
Five pre-set modes
Pair of pads
15 minute shut off.
thorax
5. The lights
across your chest
Battery Light
Pad Light
BUTTONS AND THEIR FUNCTIONS
calves of both legs
Mode Light
5
soles of both feet
Mode/Intensity Button_
skin that lack normal sensation.
Cord.Jack------ ---------- ---__,
before using this unit.
in order to confi r m that it is advisable for you to use this Electronic Nerve Stimulator
before using this unit.
by the manufacturer to avoid damage to
are
on the part of the body you wish to treat or the therapy may
such as a belt buckle, necklace, or other metal worn under clothing.
or put pads on top of each other. It may weaken or stop therapy, or the unit may stop working.
and cause gel to be removed when separating.
This may cause a skin irritation or
because the gel may get damaged and it won't stick or function properly.
provided (Refer to [GET STARTED WITH YOUR THERAPY -STEP
to the pads or to your skin because it will disrupt the pads from functioning properly. The
or you feel discomfort,
explode.
except for what it is
LL
such as a
environmental pollution.
shorten the battery life.
&. During therapy, do not remove the battery cover and do not
touch the battery terminals.
LL Do not use rechargeable batteries.
than
&
affect the safety and effectiveness of electrical stimulation.
during treatment.
&.
LL
could result in discomfort or skin
burns.
LL Detach the pads before replacing the batteries.
due to the electrical stimulation or electrical conductive medium (gel).
The unit contains small pieces that may be swallowed. Immediately contact your
you experience adverse reactions from the unit.
beneath the stimulation electrodes applied to your skin.
during or following the application of electrical stimulation near your
Pad Holder
Batteries*
(1 standard)
(2
Back
* The supplied batteries are for
trial use only, they may have a
shorter life.
(EB
Instruction Manual
Quick Start Guide / Pad
Placement Guide
Long L1
Pads
· r e
(Arm/Shoulder, Lower Back, Leg/Foot, Knead, Steady).
((W[GET STARTED WITH YOUR THERAPY -STEP 3 -
SELECT INTENSITY LEVEL (I LOW - 10 HIGH)])
:
:
Press once for "ON" and again for
' ,(�': {i'·
If the pad dislodges Pad Light will blink .
"-"'
�--'P_,,
((W[TROUBLESHOOTING])
-
/
.:--::
Cl
Light
. 1
I
fth b
e attery 1s ow, Battery L1g I w1 1g t up.
After choosing the mode: Press Set/Start to select the mode.
((W [GET STARTED WITH YOUR THERAPY -STEP 2 -
SELECT I OF 5 MODES])
IMPORTANT SAFETY PRECAUTIONS AND WARNINGS
pain, consult with your physician
more than five days,
consult your physician
consult with your physician
You MUST USE TWO PADS at the same time.
connected well and the pads are fixed
put the pads only on the skin or on the plastic pad holder provided.
BATTERY
Lt. Do not throw the batteries into a fire.
Dispose of the unit, batteries, and components according
to applicable legal regulations.
Ideal temperature for
Lt. Do not mix alkaline and manganese batteries
PADS
Use only pads recommended by the manufacturer
The electrical performance characteristics of pads may
affect the safety and effectiveness
hold the plug and
Applying pads incorrectly
KNOW YOUR UNIT
(EB ( :E
01TIR0fl
-•lo:-$tlmul-
6�
.
.
lrllll•r-1,11..,.
-�
c•
� �
Intensity Light
"
I
Power Button
f
TI
: :
f
)
.
[ (�;
(§;;,
;;
1IL P�
Pad Light
.
r/ /
om"""
d
'
1/
Battery
Set/Start Button
stop using the unit and
Remove the battery cover using a hard object (such as the tip of
a pen, thin screwdriver).
Insert batteries. Make sure the Ee 8 signs correspond when
inserting batteries.
Reinstall the battery cover.
Attach the electrode cord plug to the bottom of the main unit.
The batteries may
Unlawful disposal may cause
as this will
to avoid
of electrical stimulation.
For optimal therapy:
Place pads on either side of the pain, not directly on the
Place pads at least 2.5 cm apart.
& MUST USE TWO PADS at a time for therapy to work.
& Do not overlap pads or put on top of each
Lt. Do not add spray, lotions or creams to skin or pads.
& Do not share pads.
physician.
or muscles in
Attach both pads on either side of the
region where you feel
pain.
batteries (LR03))
Do not use near the heart, on both sides of the thorax or across your chest because the introduction of
electrical current may cause rhythm disturbances which could be lethal.
Attach both pads on the lower back according to your pain .
Place pads on muscle of back, not on spine, for optimal therapy.
"OFF".
&
Attach both pads on
• h
•n r h
Attach both pads on the calf where
either side of the area
you feel
pain.
Pads should not be placed
with pain.
simultaneously on the calves of
both legs.
Before using your unit, inspect these things.
1. Cord is not broken.
2. Pad gel is not damaged.
3. Electrode cord connection is not broken.
4. Unit is intact and working.
5. There is no battery leakage.
STEP 1 - INSERT BATTERIES
2
3
STEP 2 - ATTACH ELECTRODE CORD
TO THE MAIN UNIT
GET STARTED WITH YOUR THERAPY
(USE FOR A MAXIMUM OF 30 MINUTES PER SESSION)
STEP 1 - PAD PLACEMENT
Lt.
Lt.
ARM
"AAN'
A
e(�
I �
e/�
LOWER BACK
LEG
(HIP
THIGH)
A
For the fi r st time, take the pads out of the sealed package.
LL
&
Pads will not stick if the skin has too much hair. We recommend
shaving the area for effective treatment.
Di�,nl <ho pl�ti, film b
s <ho do� P"'l r agi" 8 .
Remove plastic fi l m and put sticky
side of pads on either side of the
pad holder.
pain.
other.
Attach both pads on the shoulder according to your pain.
Attach one pad on the front and on the back of your shoulder.
Attach one pad below and
above the region in pain,
both on same side.
Outside
Attach pads on the left for pain on the outside of
your ankle/foot. Attach the pads on the right for
pain on the inside of your ankle/foot.
Do not put the pads on the bottom of both feet
at the same time.
ASSEMBLY STEPS
STEP 3 - SNAP EITHER ELECTRODE CORD TO
EACH OF THE PADS
Do NOT turn unit on,
until pads are on your
skin.
You MUST USE BOTH
PADS or stimulation will
not work.
NOTE:
STEP 4 - REMOVE AND DISCARD PLASTIC FILM
FROM PADS
For the first time, remove the clear plastic film from the back
of the pad.
STORING PADS ON PAD HOLDER
i
--Pad facing down
-Pad holder
--Pad sticky side up
11
� ,.
J l
�'�
SHOULDER
SHOULDER
LOWER BACK
LEG
FOOT
(CALF)
(ANKLE)
A

Advertisement

loading

Summary of Contents for Omron HV-F013

  • Page 1 Children should be supervised to ensure they do not use the equipment as a toy. (Arm/Shoulder, Lower Back, Leg/Foot, Knead, Steady). 2. Ten intensity levels (1 low to 10 high). (durable, re-usable, washable, up to 150 uses). (Only use OMRON manufactured pads or cord with this unit). 4. Automatic or any area of the Attach both pads on the lower back according to your pain .