Controller Specifications - Agilent Technologies Turbo-V 750/850-AG User Manual

Rack controller
Table of Contents

Advertisement

17
Technical Information

Controller Specifications

Controller Specifications
Tab. 2
Input:
Voltage
Frequency
Power
Output for pump:
Power during pump ramp-up
Power normal (pump water cooled)
Power normal (pump air cooled)
Operating temperature
Storage temperature
Fuse:
Mains
Primary pump control Output
Active Gauge Input
Serial communication (T-Plus kit)
Power cable
Pump cable
In compliance with norms
Max altitude
Protection category
Internal use only
Weight (both models)
Installation category
Pollution degree
212/304
100 ÷ 240 Vac (±10 %), 1-phase
50 to 60 Hz
450 VA maximum
320 W maximum
300 W maximum
200 W maximum
5 °C to +45 °C
-20 °C to +70 °C
2 x T6.3 A (slow blow) 250 V 5x20 fuses
2 configurable ports
48 Vdc (±10 %), 250 mA maximum
2 ports compatible with all Agilent gauge models
RS232 cable with a 9-pin D type male connector and a 9-
pin D type female connector, and T-Plus software
(optional)
With European or NEMA plug 3 meters long (optional)
5 meters long fixed cable
EN 61010-1 (2001)
EN 61326-1 (2006)
Emission level : Class B for Residential Application
Immunity level : for Industrial Application
2000 m
IP 20
6.3 kg (13.9 lbs)
II
2
Turbo-V 750/850-AG Rack Controller User Manual / 87-901-023-01 (B)