Clean the spray head with a respective solvent if the sprayer is not used for a longer
time

4.2 Decontamination

If the instrument needs to be decontaminated, follow the respective safety
regulations/guidelines if applicable.

4.3 Troubleshooting

In case of insufficient performance check:
Loading state of the battery
Level of solution in the reagent bottle
Clogged nozzle: clean carefully, using a fine needle. Make a test spray prior to next
application. If this does not help, exchange the spray head.

5 Technical data

Charger unit
Mains voltage: 100 - 240 Volt / 50 / 60 Hz
Power consumption: approx. 5 Wh
Pump unit
Operating voltage: 4.8 V DC
Inductive charge: 45 - 50 mA / 14 - 16 hours
Avoid contact of water or organic solvents with the sprayer or the charging unit.
The TLC Sprayer meets all relevant safety regulations as EEC safety regulation 82/499; VDE
(Assoc. of German Engineers) spark protection 0871-B.
Telephone
CAMAG Scientific Inc. (USA) · 515 Cornelius Harnett Drive · Wilmington, NC 28401
Instruction manual, May-18
CAMAG (Switzerland) · Sonnenmattstrasse 11 · CH-4132 Muttenz 1
Telephone
41 61 467 34 34 · Fax
+
CAMAG (Germany) · Bismarckstraße 27-29 · D-12169 Berlin
49 30 516 55 50 · Fax
+
Telephone 800 334 3909 · Fax 910 343 1834 · E-Mail:
41 61 461 07 02 · E-Mail:
+
49 30 795 70 73 · E-Mail:
infoberlin@camag.com
+
tlc@camag.com
www.camag.com
TLC Sprayer
info@camag.com
7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hptlc

Table of Contents