HP 431B Calibration Procedure page 14

Power meter
Hide thumbs Also See for 431B:
Table of Contents

Advertisement

TB 9-6625-086-35
(4) Set the TI MOUNT RES switch to agree with the thermistor mount (A8) being
used.
(5) Set the TI RANGE switch to .01 MW and null and zero the meter.
(6) Adjust dc power supply (A2) for a full-scale (.01 mw) indication on the TI. If dc
voltmeter (A3) indication is not within the limits specified in table 5, perform b below.
Figure 4. Range and tracking check without power meter calibrator.
Table 5. Range and tracking Performance Check (Without Power Meter Calibrator)
Test
Test
Instrument
Instrument
Switch
Indication
Settings
.01
.01
.03
.03
.1
.1
.3
.3
1
1
10
Dc Voltmeter Indications (vdc) (Readings
for Model 431B are in Parenthesis)
200-Ohm Mount
Min
Max
.438
.456
(.434)
(.460)
.763
.787
(.752)
(.798)
1.40
1.42
(1.37)
(1.45)
2.43
2.47
(2.38)
(2.52)
4.43
4.51
(4.34)
(4.60)
100-Ohm Mount
Min
Max
.619
.645
(.613)
(.651)
1.08
1.12
(1.06)
(1.13)
1.98
2.02
(1.94)
(2.06)
3.43
3.49
(3.36)
(3.56)
6.26
6.38
(6.13)
(6.51)
Adjustments for Model
431C (fig. 3)
200-Ohm
100-Ohm
Mount
Mount
R14
R1
R13
R2
R12
R3
R11
R4
R10
R5