HP 431B Calibration Procedure page 12

Power meter
Hide thumbs Also See for 431B:
Table of Contents

Advertisement

TB 9-6625-086-35
(6) Position the controls of the power meter calibrator as follows:
(a) RANGE (MW) switch to .01.
(b) FUNCTION switch to CAL.
After each change in position of TI RANGE switch, turn the
FUNCTION switch of power meter calibrator to the CURRENT
OFF position and adjust the TI zero controls for a zero
indication.
(7) Turn the TI and power meter calibrator RANGE switches to the settings listed
in table 4. If the indications are not within limits specified, perform b below.
Table 4. Range and Trucking Performance Check (with Power Meter Calibrator)
Power Meter
Test Instrument
Calibrator
RANGE switch
RANGE (MW)
settings
switch settings
.01
.03
.1
.3
1
3
10
10
1
10
10
10
8
Test Instrument Indications (mw)
Min
.01
0.0098
(0.0097)
.03
0.0295
(0.0291)
.1
0.099
(0.097)
.3
0.297
(0.291)
1
0.990
(0.970)
3
2.970
(2.910)
10
9.900
(9.700)
8
7.9
(7.7)
6
5.9
(5.7)
4
3.9
(3.7)
2
1.9
(1.7)
NOTE
(Readings for Model 431B
are in parenthesis)
Max
0.0102
(0.0103)
0.0305
(0.0309)
0.101
(0.103)
0.303
(0.309)
1.010
(1.030)
3.030
(3.090)
10.100
(10.300)
8.1
(8.3)
6.1
(6.3)
4.1
(4.3)
2.1
(2.3)
Adjustments for Model
431C (fig. 3)
200-Ohm
100-Ohm
Mount
Mount
R14
R1
R13
R2
R12
R3
R11
R4
R10
R5
R9
R6
R8
R7