Error Message
Supply Pressure
Too Low
Value out of range
Function not
available
Illegal Operation
PMM uninstalled
PMM unsupported
46
Table 6. Error Messages (cont.)
Explanation
Failed to create supply
pressure for the current
setpoint.
Value input is out of
range.
Selected function is not
available for the current
state.
Illegal operation for the
current state.
Pressure Measurement
Module is not
connected.
Pressure Measurement
Module is not supported.
Potential Root Cause and Solution
1. The pressure volume is too large,
reduce the volume and try again.
2. The leakage of pressure volume is
too large. Test the leakage and try
again.
3. Ambient pressure is lower than the
negative gauge setpoint. Change the
negative gauge setpoint to a
reasonable value at current ambient
pressure.
4. The unit needs service.
Try to configure a valid value.
NA
NA
Make sure that the Pressure
Measurement Module in installed.
1. Range of Pressure Measurement
Module is greater than the range of
the Product.
2. Replace with supported Pressure
Measurement Module.