HP 8340A Service Manual page 28

Synthesized sweeper 10 mhz to 26.5 ghz
Hide thumbs Also See for 8340A:
Table of Contents

Advertisement

No.
Name
82
LOCK WORD
ENABLE
83
84
85
86
87
88
89
90
1
9
92
93
94
95
96
97
MODEL#
98
99
CHECKSUM
Model 8340A - Service
Table
BA-3.
Calibration Constants (7 of 7)
Function
Enables lockword function which is
used to allow access to calibration
constants only upon input of the correct
password (lockword).
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
Instrument model number.
Serves as check point for Cal Constant
Accuracy. The sum of Cal Constants
through 98 is calculated, the sum is
complemented, and the result is stored
as the CHECKSUM.
Scans by HB9HCA and HB9FSX
Range
Significance
(Counts)
Disabled/
0/-23,875
1,
0,
99
0=8340A
1 =8341A
99=8340A-H02
-
1
Enabled
-
8-27