The connection of a radiating antenna is an illegal act in many
countries. Only connect the Product Microwave output equipment or
transmission lines designed to prevent RF leakage at the level and
frequency of the Product output.
Additional notes regarding good practice when sourcing and measuring high- and
low-level signals are given at the end of this chapter.
To connect a Leveling Head to a UUT:
1. Read and observe all of the preceding Cautions and Warnings.
2. Remove the plastic connector protection caps form the cable-end connectors
3. Connect the N-connector on the Leveling Head to the input of the UUT.
4. Torque the N-connector to 1.00 Nm (9 in-lb) using an N-connector torque
3-40
The 96040A-xx Heads are fed via very high-grade
flexible coaxial transmission line. As with any coaxial
line, deformation of sidewalls or abrupt bending can
degrade performance. Take care to prevent mechanical
stress or tight bend radius <60 mm (2.4 in).
The maximum output level of the 96040A is unusually
high (+24 dBm into 50 Ω and +18 dBm into 75 Ω). Many
RF loads, active and passive, could be damaged by this
power level. Be careful not to exceed the maximum
ratings of the any connected load
To prevent personal injury and to prevent the leakage
or transmission of an RF signal, never connect the
Product output (the output from a Leveling Head) to a
radiating antenna of any kind. Such a transmission
would be hazardous to personnel and may impair the
SAFE operation of equipment, and communication and
navigation systems.
and save them for future use.
wrench.
The torque wrench is available as an accessory; see Chapter 1, Options and
Accessories.