Download Print this page
BW Technologies GasAlertClip Instruction Sheet

BW Technologies GasAlertClip Instruction Sheet

2 year gas detector

Advertisement

Quick Links

BW TECHNOLOGIES
2 Year Gas Detector
Instruction Sheet
Introduction
The GasAlertClip gas detector ("the detector") warns of hazardous gas at
levels above (below for oxygen deficiency) a factory set alarm setpoint. The
detector is a personal safety device that stores and transmits gas alarm
event data. It is your responsibility to respond properly to the alarm.
Models with internal vibrator
alarm and event logging
GA24A-H-EL
hydrogen sulfide (H
GA24A-M-EL
carbon monoxide (CO)
GA24A-S-EL
GA24A-X-EL
GA-TC-1 Test Cap Accessory
The test cap accessory is designed to fit over the sensor when you test the
detector's response to the target gas.
GPR-PRINTER Cordless Thermal Printer Accessory
The cordless printer accepts and prints stored gas alarm event data.
GA-TS02 Automatic Test Station Accessory
The test station is designed to automatically test response to gas.
D2087/2
BW1007-27-01-12.375X11.75-5187-2
© November 2002 BW Technologies. All rights reserved.
aSafety Information - Read First
Warning: Substitution of components may impair intrinsic safety.
GasAlertClip
Do not activate the detector after the date on the package.
This product is a gas detector, not a measurement device.
Perform a self-test each day prior to use.
Make sure the sensor grill is free of dirt and debris.
Make sure the sensor grill is not covered.
Periodically test the sensor's response to gas by exposing the
detector to a target gas concentration that exceeds the low alarm
setpoint. Manually verify that the audible and visual alarms are
activated.
Periodically calibrate the GasAlertClip O
Gas Monitored
This instrument contains a lithium battery. Do not mix with the solid
S)
2
waste stream. Spent batteries should be disposed of by a qualified
recycler or hazardous materials handler.
sulfur dioxide (SO
)
2
International Symbols
oxygen (O
)
2
Symbol
h
ATEX
g
X
.
2
c
Note:
Meaning
Classified to both U.S. and Canadian Safety
standards by Underwriter's Laboratories, Inc.
Conforms to European Directive 94/9/EC – ATEX 95
European Explosion Protection
Conforms to European Union directives
Activating the Detector
Press C and hold for 5 seconds. The detector emits one beep, one flash
and one vibration*. Then the detector sets the life-remaining clock, performs
a self-test and displays the alarm setpoints. Once activated, the detector
cannot be turned off.
u
If the detector passes the self-test,
shows on the display and the unit is
ready for use. Instructions describing what to do in the event of a self-test
failure appear in the section "Self-Test Fail."
*Vibrator models only.
A
Infrared download port*
B
Audible alarm
C
Visual alarm
D
LCD display
E
Clip
F
Test button C
G
Sensor and sensor grill
*EL models only.

Advertisement

loading

Summary of Contents for BW Technologies GasAlertClip

  • Page 1 Introduction detector to a target gas concentration that exceeds the low alarm The GasAlertClip gas detector (“the detector”) warns of hazardous gas at setpoint. Manually verify that the audible and visual alarms are levels above (below for oxygen deficiency) a factory set alarm setpoint. The activated.
  • Page 2 Maximum Gas Exposure The detector records the maximum gas exposure under alarm condition, Model Low Alarm Setpoint High Alarm Setpoint and begins calculating the number of hours since the maximum exposure GasAlertClip O 19.5% 23.5% occurred. GasAlertClip CO 35 ppm...
  • Page 3 Calibrate the detector only in a normal LCD reads Test, display the alarm setpoints, and display the maximum atmosphere (approximately 20.9 % O ). To calibrate the GasAlertClip O exposure to gas if applicable. •...
  • Page 4 Up to 10 events encountered LIMITED WARRANTY & LIMITATION OF LIABILITY minutes of alarm time/day BW Technologies Ltd. (BW) warrants this product to be free from defects in material and Data transmission method Via infrared port to thermal printer workmanship under normal use and service for a period of two years, beginning on the date...

This manual is also suitable for: