3.9

Inspecting the leak test function

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
To prevent the endoscope from being damaged by water leakage, make sure to
inspect the leak test function before use. If any irregularity, such as no air supply
from the metal connector of the leak test air tube is found, do not use the
equipment and contact Olympus. Endoscope leak testing can be performed
outside the OER-Pro, as per existing labeling, or can be performed in the
OER-Pro. When the leak test is performed with the OER-Pro, follow the
procedure in Section 4.7, "Leak test" on page 102.
Check Required item
Leak test air tube (MAJ-821)
Table 3.7
Make sure that there are no cracks, breaks, fissures,
scratches, or stains on the leak test air tube. Using an
abnormal or damaged leak test air tube may result in
inaccurate inspection of leak test function.
When connecting the leak test air tube, ensure that the tube
connector is fully and properly attached. Improper connection
will result in inaccurate inspection of leak test function.
Prevent water from getting into the leak test air tube.
Otherwise, the water inside the tube may damage the
endoscope when the leak test is performed next time.
If water does not start filling the reprocessing basin about
10 seconds after the leak test has started, press the
STOP/RESET button to stop testing. Check whether the
water faucet is open. If not, open it and start leak test again.
Chapter 3 Inspection Before Use
55