Audio Measurements - Agilent Technologies E1852B Operating Manual

Bluetooth test set
Table of Contents

Advertisement

E1852-90007.book Page 85 Monday, February 4, 2002 2:09 PM

Audio Measurements

E1852B Operating Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Audio Measurements
The Audio In and Audio Out rear-panel connectors can be used to
transport an audio frequency signal to and from the DUT using a
Normal SCO Link.
Connections
Using the audio functions, you can send a signal from a known,
calibrated audio source via the test set and use an audio analyzer to
make measurements such as SINAD, frequency response, gain, noise,
and distortion on the recovered audio signal from the DUT. Similarly,
the reverse signal path may be tested. The audio signal is injected at
the DUT, recovered from the test set and measured. Figure 52 shows
some possible signal path configurations.
Audio In
Audio Signal
Generator / Mic
Audio Signal
Analyzer / Spkr
Figure 52
Audio Test Connections
Making Measurements
RF
DUT
Audio Out
Audio Signal
Analyzer / Spkr
Audio Signal
Generator / Mic
85