Table of Contents

Advertisement

Cooling Requirements
For 15 °C rise 4.5 liters/second 0.9 mm H

Timing

Glitch-free timing changes
You can sweep your timing values without danger of spurious pulses or drop-
outs that could cause measurement errors. This applies to continuous mode with
timing values < 100 ms (frequency: > 10 Hz), and consecutive values between
one-half and twice the previous value.
Period
Period can also be entered as frequency.
Period
Range:
Resolution:
Accuracy:
(add ± 100 ps in burst mode)
RMS-jitter:
Frequency range:
There are 2 period generation sources available:
• startable oscillator (variable frequency oscillator VFO) with fixed latency
from trigger input to outputs.
• high-accuracy frequency generator (PLL) .
Agilent E8305A Specifications
O.
2
Value
4 ns to 1000 s
3.5 digits, 5 ps best case for VFO
4 digits, 1 ps best case for PLL
PLL: 0.01%
VFO: 0.5% after selfcal, typical
5% w/o selfcal
PLL: 0.001% + 15 ps
VFO: 0.01% + 15 ps
1.00 mHz to 250 MHz
Specifications
93

Advertisement

Table of Contents
loading

Table of Contents