Maintenance Schedule - Meyra 1.620 Operating Manual

Electric wheelchair
Hide thumbs Also See for 1.620:
Table of Contents

Advertisement

Maintenance schedule

WHEN
Before starting out
Especially before driv-
ing in the dark
Every 2 weeks
(depending
on
tance covered)
Every 6 -8 months
(depending
on
tance covered)
WHAT
General
Test for faultless operation.
Checking the magnet-
ic brake
Move the selection lever
for the drive/push mode
into the drive mode posi-
tion on both sides.
Lighting
Check the lighting equip-
ment and reflectors for
flawless functioning.
Check air pressure of
dis-
the tyres
Tyre filling pressure:
2.5 bar = 36 psi
Adjustment screws
Screws and nuts are to be
checked for tight fit.
Wheel attachments
dis-
Wheel nuts or screws are
to be checked for tight fit.
REMARK
Carry out test yourself or with
a helper.
Carry out test yourself or with
a helper.
If the electric wheelchair can
be pushed, have the brakes
repaired immediately by the
specialist workshop.
– Danger of accident!
Carry out test yourself or with
a helper.
Carry out test yourself or with
a helper.
Use a tyre gauge.
Carry out test yourself or with
a helper.
Retighten the loosened ad-
justment screws.
Contact specialist workshop
upon demand.
Do it yourself or with the aid
of a helper.
Securely tighten any loosened
wheel nuts or screws and
retighten again after 10 oper-
ating hours or resp. 50 km.
Contact specialist workshop
upon demand.
41

Advertisement

Table of Contents
loading