GasAlertMicro 5/PID/IR
User Manual
L
J
If enabled, during an alarm condition the Latch (latching
alarms) option causes the low and high gas alarms (audible,
visual, and vibrator) to persist until the alarm is acknowledged
by pressing C and the gas concentration is below the low
alarm setpoint. The peak concentrations display continually
until the alarm condition no longer exists. Local regulations
may require the Latch Alarms option be enabled.