Calibration Adjustment
1. Using the equipment connections shown in Figure 2, set the 5520A output for 2 A, 45 Hz.
2. Referring to Figure 3, slide the rubber boot on the UUT handle back to access the adjustment point,
VR1.
3. Using a small flat-head screwdriver, adjust VR1 for a meter reading of 99.9 mV to 100.1 mV.
User-Replaceable Parts
The Current Clamp contains no user-replaceable parts. The following user documentation is available for
this product.
i400s Instruction Sheet
i400s Calibration Information
(this document)
1.800.561.8187
Description
www.
VR1
Figure 3. i400s Calibration Point
Part No
2282671
2414327
.com
AC Current Clamp
Calibration Adjustment