Sourcing mA
Use source mode whenever you need to supply current into a
passive circuit. Insert the test leads into the terminals as shown
below.
A path must exist for current to flow between the output
+ and −mA terminals, otherwise the display will indicate
an overload (OL) when you set an output value.
Source
Press to step output
up/down in 25 %
steps.
Press to toggle
Source or Simulate.
www.
English Instruction Sheet
Note
or
705
LOOP CALIBRATOR
Press to scroll
up/down 0.01 mA.
Hold down to scroll
faster.
For 0.001 mA steps,
see "Pushbutton
Functions."
MEASURE
OFF
SOURCE
VDC
SIMULATE
Loop Power
mA
SpanCheck
Press simultaneously
+
COM
once for slow ramp
twice for fast ramp
three times for
stepped ramp.