717 Verification; Pressure Verification - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

Table 61. 715 DC Voltage Measure Verification (0.0000 mV to 200.0000 mV)
Fluke 5520A
0.0000 mV
50.0000 mV
150.0000 mV
6. Disconnect the test leads from 715 Calibrator and turn the green  key off.
The 715 verification test is complete.

717 Verification

Pressure Verification

1. Carefully attach the pressure fitting of the deadweight tester to the pressure port of
the 717 30G Calibrator.
Use Plenty of TEFLON tape when attaching pressure fitting.
The display should read 0.00 PSI with the deadweight tester opened up to ambient air. If
not, push the
2. Set up the deadweight tester for the sequence of PSI inputs from Table 62 to be
injected into the pressure port of the 717 Calibrator.
In Table 62, Vacuum is only specified for V1.3 and later. To display the
firmware version:
1. Turn off the 717.
2. Push and hold the center bottom button as you push
3. Ensure the pressure has stabilized at each input before verifying the display reading.
Note
key until display reads 0.00 PSI.
Note
Fluke 715
-0.02 mV to + 0.02 mV
49.98 mV to 50.02 mV
149.96 mV to 150.04 mV
to turn power on.