Adjust 24 mA Input

1.
Set the 5522A to output 24 mA.
2.
Push  on the 712B to enter 24 mA input calibration mode.
Wait approximately 10 seconds for a stable output of the 5522A.
3.
Confirm the reading on the 712B.

Adjust External Trigger for Ohms Source

Before the calibration, set the 5522A to mA (dc) source output mode, and use
one set of banana plug patch cords to connect the 712B and the 5522A, as
shown in Figure 14.
Fluke 712B
Figure 14. Adjust External Trigger for Ohms Source
RTD/Thermocouple Calibrator
712B Calibration Adjustment
hwp15.jpg
Fluke 5522A
5522A
CALIBRATOR