2 Overview

Through the broading market investigation, ous company assembles a lot of talents in the
following majors: the appearance design of the product, structure design, transmission structure
design, and circuit design, they produced the equipment successfully after their deep research
and reduplicate tests. This equipment adopts the thermal recording mode the key point of thermal
recording technique is the core component---thermal recording probe. The thermal recording
probe is the high-tech component with advanced component integrated technology, integrating a
large amount of heating elements (8points/millimeter) and its controlled circuit based on the
ceramic basal body. It doesn't have the false difference which caused by non- linearity, damping,
and incongruence of pen warmth of the traditional hot pen ECG. In the frequency response
which is the important index of affecting the descriptive authenticity, The equipment can reach
0.05-1 SOHz frequency response which is the world level, also the equipment has advanced
digital baseline drifting filter, Alternating Current filter, and Myo- electricity disturbance filter.
The sampling rate of lead signal is 800Hz.the closure frequency of low pass filter machine with
guards against mixes folds is (-3dB) 220Hz. The sampling rate is higher than the standard of not
lower than 500Hz which made by American AHA Association and European CSE Association.

2.1 Features

>
Operation of the product is simple and convenient.
>
The product is small in volume, light in weight (total weight is about 800g including
batteries) and convenient in carrying.
>
The instrument can be used under AC&DC, with the lithium rechargeable battery inside,
and it will record eighty ECG waveform persistently under the best DC conditions.
>
Voltage indication will be displayed when power on.
>
Classification: CF
>
EMC complies with the requirements ofIEC60601-l-2:2001 (refer to the appendix 2)

2.2 Major applications and scope of application

This product is single channel, 12 leads electrocardiograph, can be widely applied in sampling
ECG under different circumstances such as hospital consultation, doctor's diagnosis, social
medical organizations etc. It can implement real time continuous records of clear and exact
single-channel ECG waveform using thermosensitive printer at the same time. Waveforms also
can be frozen at any time. It has manual and automatic modes to be chosen and Chinese/ English
operation interface. It is easy to be used.