A 803 Holder For Transmission Instruments; Installation; Preparing Transmission Instruments - Miele A 105 Operating Instructions Manual

Upper injector basket
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 31
en - Areas of application

Installation

Preparing
transmission
instruments
40

A 803 holder for transmission instruments

An A 803 holder has to be retro-fitted for reprocessing dental trans-
mission instruments. This holder is suitable for almost all standard
transmission instruments.
Remove an injector nozzle from the injector manifold and replace it
with an A 803 holder.
The suds solution has to be filtered when using an A 803 holder. To
do so an A 800 tubular filter has be fitted inside the injector mani-
fold.
The tubular filter is supplied with its own instructions.
In particular please observe the instructions on regular cleaning of
the tubular filter.
Carry out a visual check of all internal sections for any blockages.
Fit the transmission instrument into the holder. The instrument is
then held in position by the spring clamp.
After reprocessing pull the transmission instrument out of its holder.
The spring clamp will automatically spring back into position.
After processing dry internal sections of transmission instruments
with sterilised compressed air and then sterilise as required follow-
ing manufacturer's instructions and national health and safety regu-
lations.
Before using transmission instruments again following repro-
cessing, a function check must be carried out, e.g. by spraying into
a basin, to ensure they are clear.