CYANSmart
Service Manual
For a clear and precise diagnose
ISO 13485-2016
www.diagnostics.be • Belgium • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
Layout 20181024

Advertisement

Table of Contents
loading

Summary of Contents for Cypress Diagnostics CYANSmart

  • Page 1 CY009 CYANSmart Service Manual For a clear and precise diagnose ISO 13485-2016 www.diagnostics.be • Belgium • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be Layout 20181024...
  • Page 3: Table Of Contents

    Verification of the aspiration switch 4.4.10 Replacement of the aspiration switch 4.4.11 Aspirate format CYANSmart Service Service Manual, version 20191121 (8.0) Page 1 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 4 SAFETY INFORMATION ........................... 65 INSTALLATION FORM MAINTENANCE FORM REPAIR FORM SPARE PARTS CYANSmart Service Service Manual, version 20191121 (8.0) Page 2 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 5: Introduction

    1 INTRODUCTION 1.1 S YSTEM DESCRIPTION The CYANSmart is an easy to operate semi-automatic biochemical analyzer. It is a microcomputer-based independent unit, controlling several systems: built-in thermal printer, touch screen, optical detection system with flowcell, incubator and aspiration system. ②...
  • Page 6: Echnical And Operative Specifications

    Relative humidity: 30 % - 70%   Air pressure: 860 hPa – 1060 hPa CYANSmart Service Service Manual, version 20191121 (8.0) Page 4 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 7: Installation

    CY009 – CYANSmart – Service Manual 2 INSTALLATION The CYANSmart is an easy to operate instrument with integrated user-friendly software. No additional computers or instruments are needed for good operations. In order to fully guarantee the performance of the CYANSmart, it is extremely important that it is correctly installed.
  • Page 8: Maintenance

    CYANSmart Service Service Manual, version 20191121 (8.0) Page 6 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 9: Cleaning

    Open the filter cover and blow the dust off. OTES The instrument cover should only be removed when the CYANSmart is unplugged.  We strongly advise to leave this procedure to the technical assistance of the service engineer of your ...
  • Page 10: Replacement Of Specific Parts

    In case of leakage, obstruction or strong coloration Service engineer CYANSmart Service Service Manual, version 20191121 (8.0) Page 8 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 11: Technical Control

    The touch screen and the printer with separate board are connected to the mainboard but integrated in the instrument cover. CYANSmart Service Service Manual, version 20191121 (8.0) Page 9 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 12: Main Control Board

    The following sections provide more explanation on the individual functions. CYANSmart Service Service Manual, version 20191121 (8.0) Page 10 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 13: Voltage Check Points On The Mainboard

    If the voltage is not correct, please proceed to the verification of the power supply (see 2). CYANSmart Service Service Manual, version 20191121 (8.0) Page 11 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 14: Replacement Of The Mainboard

    Temperature sensor Fan of optical system ④ ⑥ ⑦ ③ CYANSmart Service Service Manual, version 20191121 (8.0) Page 12 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 15: Lamp

    (2) and the main control board. All filters of the CYANSmart except the 340 nm, have a pre-filter installed before the filter. This is a non- selective absorbent circle with a central hole with dimensions from 0.2 mm to 2.5 mm. The pre-filter will decrease the light intensity of the light source before it passes the filter.
  • Page 16: Pre-Amplifier Board

    3. Pressing the “Read” button, the instrument will perform a new reading and display the results. CYANSmart Service Service Manual, version 20191121 (8.0) Page 14 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 17 2. Verification of the movement of the filter wheel (Proceed to 4.3.7 and 4.3.8 and 4.3.9 and 4.3.10). CYANSmart Service Service Manual, version 20191121 (8.0) Page 15 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 18: Cleaning And/Or Replacement Of The Flowcell

    Insert the flowcell completely to the bottom of the support. Make sure nothing is blocking the insertion! Secure its position with the designated golden plate. CYANSmart Service Service Manual, version 20191121 (8.0) Page 16 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 19: Verification Of The Filter Wheel Home Sensor

    546 nm  Light green 578 nm  Yellow 620 nm  Red CYANSmart Service Service Manual, version 20191121 (8.0) Page 17 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 20: Cleaning And/Or Replacement Of The Filter Wheel Home Sensor

    (see 4.3.15). CYANSmart Service Service Manual, version 20191121 (8.0) Page 18 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 21: Verification Of The Filter Wheel Motor

    8. Remove the 2 screws attaching the home sensor and carefully take it out. CYANSmart Service Service Manual, version 20191121 (8.0) Page 19 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 22 12. Unscrew the 4 hex/allen screws that attach the filter wheel motor to the optical system, and take it out. Back view CYANSmart Service Service Manual, version 20191121 (8.0) Page 20 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 23: Cleaning And/Or Replacement Of Filters

    (arrow pointing away from the lamp and towards the flowcell/pre-amplifier board). CYANSmart Service Service Manual, version 20191121 (8.0) Page 21 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 24: Verification Of The Voltage To The Lamp

    Test point 8 compared to the grounding of test point 5 (see 4.2.2 or on picture below left) or a screw. The measured voltage should be 6V (DC) in normal mode. If the CYANSmart is in standby mode the measured voltage will be 3V (DC).
  • Page 25: Replacement Of The Lamp

    √ c) Dust and dirt influence the gain/AD/offset values. So before changing the lamp or the pre- amplifier board, make sure the CYANSmart is clean and dust free, both the hydraulic system and the filters (see 3.3.1 and 3.3.2). 4.3.13.1 Installing the lamp Procedure: 1.
  • Page 26 (reuse the tightening screw). Turn the screw clockwise to fix it. Connect the lamp wire to the mainboard (see 4.2.1). : After the lamp replacement, turn ON the CYANSmart and wait 30 minutes ARNING before adjusting the height of the lamp. It is necessary to stabilize the light emission.
  • Page 27 CYANSmart Service Service Manual, version 20191121 (8.0) Page 25 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 28: Replacement Of The Pre-Amplifier Board

    The capture efficacy will diminish. This will first influence the detection of low absorbance. Therefore, Cypress Diagnostics recommends to replace the pre- amplifier board every 2 or 3 years together with the replacement of the lamp.
  • Page 29 6. To install the new part, repeat the above steps in reverse order. CY009-S01 CYANSmart Service Service Manual, version 20191121 (8.0) Page 27 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 30: Hydraulic And Aspiration System

    Verify the precision/accuracy of the pipette used  (re)calibrate the pipette if necessary. CYANSmart Service Service Manual, version 20191121 (8.0) Page 28 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 31 If all the above is verified, but the pump calibration remains out of range, the pump tube or pump rubbers needs to be replaced (see below). CYANSmart Service Service Manual, version 20191121 (8.0) Page 29 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 32: Replacement Of Tubes In The Hydraulic Circuit

    It is recommended to periodically check the level in the waste container, in order to avoid overflow. CYANSmart Service Service Manual, version 20191121 (8.0) Page 30 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 33: Replacement Of Peristaltic Pump Tube

    4. If the instrument is not used for more than one week, unhook the hinged guide lock to prevent the pump tube from sticking together. CYANSmart Service Service Manual, version 20191121 (8.0) Page 31 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 34: Replacement Of Peristaltic Pump Rubbers

    3. Unhook the hinged guide lock and remove the pump tube from the pump rotor. CYANSmart Service Service Manual, version 20191121 (8.0) Page 32 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 35: Verification Of The Pump Motor

    8-10 Ω. Otherwise, the pump motor needs to be replaced. CYANSmart Service Service Manual, version 20191121 (8.0) Page 33 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 36: Replacement Of The Pump Motor With Rotor

    7. Otherwise, the aspiration switch with wire needs to be replaced. CYANSmart Service Service Manual, version 20191121 (8.0) Page 34 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 37: Replacement Of The Aspiration Switch

    If the air gap is too small, there is an increased risk for cross-contamination. CYANSmart Service Service Manual, version 20191121 (8.0) Page 35 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 38: Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium

    CYANSmart Service Service Manual, version 20191121 (8.0) Page 36 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 39: Temperature Calibration

    37°C ± 1°C. Otherwise proceed to step 9. CYANSmart Service Service Manual, version 20191121 (8.0) Page 37 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 40 “Yes” to save the changed settings, “No” to exit the menu without saving or “Cancel” to return to the “Reserved settings” menu. CYANSmart Service Service Manual, version 20191121 (8.0) Page 38 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 41: Cleaning And/Or Replacement Of The Fan Of The Optical System

    5. Use pressurized air (dry!) or a damp cloth for cleaning the fan. ① ① ② ② ① ① CYANSmart Service Service Manual, version 20191121 (8.0) Page 39 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 42: Verification Of The Peltier Element

    5. Remove the ventilator by unscrewing its 2 screws (2) and then remove the heat sink by its 2 screws (3) CYANSmart Service Service Manual, version 20191121 (8.0) Page 40 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 43: Verification Of The Heater Element Of The Incubator

    ± 1°C, the heater element needs to be verified. Procedure: CYANSmart Service Service Manual, version 20191121 (8.0) Page 41 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 44: Replacement Of The Heater Element Of The Incubator

    (=thermal grease) on the inside in order to assure efficient heat conductivity. CYANSmart Service Service Manual, version 20191121 (8.0) Page 42 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 45: Verification Of The Temperature Sensor

    CYANSmart Service Service Manual, version 20191121 (8.0) Page 43 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 46: Replacement Of The Temperature Sensor

    Test point 4 compared to the grounding of test point 5 or a screw (see 4.2.2): 15 V (DC)  CYANSmart Service Service Manual, version 20191121 (8.0) Page 44 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 47: Thermal Carbon Printer

    4.7 T HERMAL CARBON PRINTER The CYANSmart is equipped with a built-in thermal printer. The text or graph is printed on coated thermo- chromic paper. This paper is selectively heated when passing over the thermal printing head, and thereby the coating on these places turns black, producing an image.
  • Page 48: Verification Of The Voltage To The Printer On The Mainboard

    6. Unscrew the 2 screws attaching the printer board to the case (5). CYANSmart Service Service Manual, version 20191121 (8.0) Page 46 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 49: Touch Screen

    Upon touching a specific function/key/menu, this button will become darker blue (indicating its activation) and the CYANSmart will make a beeping sound. When data needs to be entered, a keyboard will appear. The keyboard will be adapted to the data that can be entered in that specific area: alphanumeric or numeric.
  • Page 50: Adjustment Of The Screen Brightness

    4. Carefully pull out the flat touch screen cable (1) from its slot on the mainboard (4). CYANSmart Service Service Manual, version 20191121 (8.0) Page 48 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 51: Functional Verification Of The Parts Of The Touch Screen

    If not correct or broken the display will get no power. CYANSmart Service Service Manual, version 20191121 (8.0) Page 49 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 52: Replacement Of The Touch Screen

    2. Insert the flat touch screen cable in the correct direction in its slot on the touch screen board. The connections should be facing the board (see 4.8.4). CYANSmart Service Service Manual, version 20191121 (8.0) Page 50 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 53: Power Supply

    ENERAL ELECTRICITY AND INSTALLATION REQUIREMENTS The power voltages suited for the CYANSmart are indicated on the back label (1). It needs to be connected to a power supply consistent with this voltage and frequency. The instrument can work at 100 – 230V AC.
  • Page 54: Replacements Of Fuses

    4. To install the new fuses, repeat the steps above (2-3) in reverse order. CYANSmart Service Service Manual, version 20191121 (8.0) Page 52 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 55: Locating Power Problem

    5. Replace the two fuses (4.9.2) 6. Reconnect power cord and switch ON (I) CYANSmart Service Service Manual, version 20191121 (8.0) Page 53 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 56: Verification Of The Fuses

    CYANSmart Service Service Manual, version 20191121 (8.0) Page 54 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 57 If the measured voltage is correct, proceed with step 4.9.6. CYANSmart Service Service Manual, version 20191121 (8.0) Page 55 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 58: Verification Of The Power Supply

    Check all the parts individually (see relevant chapter for more information). Procedure for CY009-S24B or CYANSmart produced after 2017/03/01; from S/N BS6IIxxx on 1. Turn OFF (O) the instrument and disconnect power cord. 2. Leave only the input connector (L, N and grounding) connected. Disconnect the output connecter from the power supply unit (- and +).
  • Page 59: Cleaning And/Or Replacement Of The Fan For Power Assembly

    6. To install the new part, repeat the steps (2-4) in reverse order. ① ① ① CYANSmart Service Service Manual, version 20191121 (8.0) Page 57 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 60: Validation

    6 mL 3 mL 12 mL 8 mL 4 mL CYANSmart Service Service Manual, version 20191121 (8.0) Page 58 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 61: Potassium Dichromate (Powder)

    The AD value has been executed and the results are within the acceptable ranges. Procedure 1. Turn ON the CYANSmart analyzer and wait 30 min until the working temperature has been reached and the lamp has stabilized. 2. Go to the AD zero menu and perform an AD auto zero measurement with distilled water. The obtained values need to be within acceptable ranges.
  • Page 62: Verification Of Linearity

    Please find the fill-in Excel sheet (with the correct formula) on our website in the distributor zone. CYANSmart Service Service Manual, version 20191121 (8.0) Page 60 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 63: Troubleshooting

    Preparation of the (working reagent) and the samples. The method has been programmed correctly. CYANSmart Service Service Manual, version 20191121 (8.0) Page 61 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 64 CYANSmart Service Service Manual, version 20191121 (8.0) Page 62 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 65: Echnical

    If after these verifications, the problems remain, see 4.6 for more details. CYANSmart Service Service Manual, version 20191121 (8.0) Page 63 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 66 Check the mentioned part, see the relevant chapter for more details. CYANSmart Service Service Manual, version 20191121 (8.0) Page 64 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 67: Safety Information

    Do not recap, purposely bend, cut, break, remove from disposable syringes, or otherwise manipulate  needles by hand. CYANSmart Service Service Manual, version 20191121 (8.0) Page 65 Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 68: Installation Form

    INSTALLATION FORM Please FILL IN the INSTALLATION FORM and send a COPY BACK TO CYPRESS DIAGNOSTICS Make sure to register also the instrument on the website: www.diagnostics.be Distributor (Company name) (Country, city) (E-mail) Coordinates (Website) (Tel.) (Name) Technician in charge (E-mail) (Tel.)
  • Page 69 Check location requirements □ Check electronical current requirements. If necessary, use: □ Electronic stabilizer □ Installation of all accessories on the CYANSmart analyser □ Connect the power cord to a suited power supply □ Install the peristaltic pump □ Install the flowcell, aspiration button and aspiration tube □...
  • Page 70: Maintenance Form

    Serial number Software version Last visit: (date) Next visit: (date) Laboratory technician Name Signature Maintenance engineer Name Signature Date Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 71  Connector of waste tube (CY009-S29)  Connector of pump tube (CY009-S30) Problems Fill in repair form* Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium • www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 72: Repair Form

    Software version Date of complaint (Date) Description of problem (details) Actions undertaken (details) Code Name Spare parts used Problem solved Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 73 We suggest to keep the original together with the instrument and to deliver copies to the maintenance engineer and Cypress Diagnostics. Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail:...
  • Page 74 See hydraulic diagram Replace every year CY004/9-S41 Tube from flow cell to pump See hydraulic diagram CY004/9-S44 Waste bottle Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 75 CY009-S01 Optical Assembly lamp, flowcell, heater, sensors, fan or board CY004/9-S02 Filter 340 nm CY004/9-S03 Filter 405 nm Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 76 Compatible with older instruments Cables and support included CY009-S16A Fan for power assembly (80x80) Originally installed until SN BS6KHxxx; Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 77 Fuse T2AL250V (10 pcs) 2 per instrument CY004/9-S49 Power cord CY004/9-S18 Power switch CY009-S46 Power plug connector (seat) Hydraulic diagram Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be...
  • Page 78 CY009 To all Healthcare Professionals, Thank you for your interest. We appreciate it! CYANSmart Cypress Diagnostics provides the tools and solutions that let clinical laboratories worldwide deliver clear and precise diagnostic results. We manufacture the equipment needed for handling the...
  • Page 79 Software • Wavelength selection: Automatic via 7 interference • Limited consumables required • All Cypress Diagnostics biochemistry reagents are filters: 340, 405, 492, 510, 546, 578 & 620 nm • Unique 16 digit patient ID • Extra items included on delivery (two fuses, five paper preprogrammed on delivery • Memory: 180 test methods and 6000 sample test results...