HP 608E Operating And Service Manual page 32

Vhf signal generator
Hide thumbs Also See for 608E:
Table of Contents

Advertisement

Section V
Model 608E/F
Table 5-1.
Test Equipment Required Cor Maintenance
Instrument
Critical Specifications
Recommended Model
Electronic Counter
Frequency Range: 10 MHz to 445 MHz
HP 5245L with 5253B
Accuracy: :1::0.005%
plug-in unit
Power Meter
Range: 10
/,LW
to 10 mW
HP 432A
Accuracy: :1::5% of full scale
Thermistor Mount
Frequency Range: 10 to 480 MHz
HP 478A
SWR:1.6max
Coaxial Attenuator
Attenuation:
10 dB
HP 8491A
Coaxial Attenuator
Attenuation:
20 dB
HP 8492A
Oscillator
Frequency Range: 20 Hz to 30 kHz
HP200cD
Envelope Distortion: 0.2% max
Output: 1 V rms
Output Impedance:
BOon
VTVM
Voltage Range: 0.1 mV to 300V
HP 400E
Decibel Range: -72 to +52 dB
Frequency Range: 20 Hz to 6 MHz
Distortion Analyzer
Range: 5 Hz to 600 kHz
HP 334A
Accuracy: ::1::3%
Deviation Meter
Frequency Range: 50 to 150 MHz
Measurements Corp.
Accuracy: ::1::3%
Model 140
Deviation Range: 0 to 5 kHz
Power Supply
-2 to -32 V variable
HP 6217A
OsCilloscope
"F(.r~ical
Bandwidth:
5 MHz
HP 140A with 1402A and
Vertical sensitivity: 5 mV/cm
1420A plug-in units
Adapter
None
UG-201A/U
Crystal Detector
Frequency Range: 10 MHz to 12.4 GHz
HP 423A
VSWR: 1. 2 max
Deviation Crom Square Law: <0.5 dB
over a range of at least 30 dB
Input Impedance:
son
Impedance Output: 15M, 10pf
Dummy Load
Resistance: 1000 :1::10%
See Figure 5-2
Wattage: 1/2 W
Pulse Generator
Pulse Width: 10$.lsec
HP 214A
Pulse Amplitude: 10 v peak into 2Kn
Pulse Rise and Decay: 0.02 $.lsec max
Repetition Rate: 50 to 5000 pps
Loop Antenna
Special
See Figure 5-1
Spectrum Analyzer
Freq: 400 MHz
HP 140S with 8554L and 8552A
Sensltivity: 1
IJ.
V. Absolute calibration
Plug-inS.
5-2