1551A/1552A Ex
Users Manual
CAL:USER:ZERO <temp>
CALC:AVER:CLE
CALC:AVER<n>:DATA?
CALC:CONV:TEST? <res>
FETC?
SENS:DATA:OHMS?
STAT:MEAS?
SYST:ERR?
SYST:PASS:CDIS
14
Table 4. List of All Commands (cont.)
Command
Command Explanation
Automatically calibrates the center of the range of
the product. The probe temperature must be held
at a constant, known temperature near 0 °C. The
<temp> value is the temperature in degrees
Celsius or Fahrenheit that the Product should be
displaying after it is calibrated. This command is
password protected. The Product temperature
units must match the units of the <temp> value.
Resets the minimum and maximum to the
present reading, and clears the stability trend
history.
Returns the value of a statistical calculation. The
AVER suffix, <n>, specifies the calculation type
as follows:
1
Maximum
2
Minimum
3
Trend
Maximum is returned if no suffix number is given.
Returns the calculated temperature in Celsius for
the given sensor reading, or "0.0,OL" if the result
is out of range. The parameter <res> is the
sensor reading in ohms.
Returns the last measurement in units according
to the UNIT:TEMP setting. If there is no valid
measurement available the response is "0.0,OL".
Returns the present ohms reading. The response
is expressed in ohms. If there is no valid
measurement available the response is "0.0,OL".
Reads and clears the Measurement Event
Register, indicating whether a new measurement
is available to be read. Returns "1" if a new
measurement has occurred since the previous
command, and "0" otherwise.
Returns a system error message if any are
present in the system error queue. Otherwise it
returns "0,"No error"'.
Disables access to password-protected
commands.