3c. Press the DELAY softkey. Enter the trigger
delay time measured in Section 8—
Determining Trigger Time Delay. The
delay time is entered in microseconds, for
example: 17 ms delay time = 17,000 µs.
Press the ENTER key twice. It is important
to enter an accurate delay time in order to
obtain accurate test results. The test may
fail frequently if the trigger does not occur
at the time of power transitions.
3d. Press the Previous Menu softkey to return to
the Time Response menu screen.
References
• Agilent Technologies 89725C Measurements
Personality User's Guide, pub. number 85725-
90010
• For Agilent Technologies 859XE setup and con-
figuration refer to the User's Guide, pub. num-
ber 08590-90301.