Maintenance And Storage - 3M Ranger 245 Service Manual

Blood/fluid warming system/irrigation fluid warming system for i.v. use
Hide thumbs Also See for Ranger 245:
Table of Contents

Advertisement

English
Ranger Blood and Fluid Warming Systems

Maintenance and Storage

C
leaning the
Clean the Ranger warming unit on an as-needed basis�
caution
• Do not immerse the warming unit in cleaning or sterilizing solutions� The unit is not liquid-proof�
• Do not clean the warming unit with solvents or damage to the case, label, and internal components
may result�
t
C
t
o
lean
1�
Disconnect the Ranger warming unit from the power source�
2�
Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute
bleach, or cold sterilants� Do not use abrasive materials�
3�
Wipe with a dry, soft cloth�
t
C
t
o
lean
The Ranger cleaning tool (part #90030) is intended to clean both heating plates of the warming unit�
It is not necessary to disassemble the warming unit to use the tool�
caution
• Do not insert metallic instruments in the warming unit, as damage to the heating plates may
occur�
• Do not use abrasive material or solutions to clean the heater plates�
• Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit�
• The cleaning tool provides only superficial cleaning—it does not disinfect or sterilize the interior of
the unit�
8
R
W
u
angeR
aRming
e
o
t
W
he
xteRioR
f
he
h
P
:
he
eating
lates
nit
u
:
aRming
nit
Arizant Healthcare Inc.