7. Alarm Display/Output

Alarm
Item
lamp
ON
. Measuring unit is
not connected.
. Connection has
been cut.
. Measuring unit
speed over.
Count response
speed over
Overflow
OFF
MG10-P1 / MG10-P2
Output
I/O connector RS-232C
Alarm
"Error" in the numerical value part of
output
the data communication output.
When this is output in mode 3, the
fifth character following the leading
character of the header changes to
"E".
[Example]
Mode 1
"00 Error"
Mode 2
"00NM Error"
Mode 3
"00NME Error"
Alarm
output
The second character in the
numerical value of the communication
data output changes to "F".
[Example]
Mode 1
"00 + F0.0000"
Cause
The measuring unit was
replaced while the power was
on.
The measuring unit is not
connected or connection has
been cut.
The measuring unit's spindle
has exceeded the maximum
response speed.
The count response speed has
been exceeded.
The number has more than 6
digits.
Solution
Reset.
Remove the cause
of the error, and
reset.
Reset.
Reset.
Limit input to 6
digits.
35