April 2020
1
Install
WARNING
Failure to follow safety instructions could result in serious injury or death.
Before installing this equipment, read the Essential instructions at the front
of this Quick Start Guide.
WARNING
ELECTRIC SHOCK
Failure to install covers and ground leads could result in serious injury or
death.
Install all protective equipment covers and safety ground leads after
installation.
CAUTION
Equipment damage
Damage to the transmitter may result.
Do not allow the temperature of the probe electronics to exceed 194 °F (90
°C)
1.1
Install probe
Emerson can supply a weld plate for welding to the flue gas duct for new
installations.
Procedure
1. If using the standard square weld plate
flange mounting plate, weld or bolt the plate onto the duct.
The through hole diameter in the stack or duct wall and refractory
material must be at least 2.5 in. (63.5 mm).
2. Insert probe through the opening in the mounting flange and bolt
through the probe gasket and flange.
For horizontal installations, the breather port must be oriented such
that it is facing downward.
Quick Start Guide
Quick Start Guide
(Figure
1-2) or an optional
5