When the GC becomes ready, the sampler or injector begins the injections.

Preparing sample vials

Selecting sample vials
The sampler and injector turrets use clear or amber glass sample vials with
crimp-caps or screw-caps. Use amber glass vials for light-sensitive samples.
Refer to your Agilent catalog for consumables and supplies for acceptable vial
types. Incompatible sample vials cause turret errors.
Figure 36
to be used with the sampler and injector. These dimensions do not make up a
complete set of specifications.
Microvial insert
Figure 36. Dimensions for sample vials and microvial inserts
Released: March 2004
Fill the solvent and waste bottles. See
bottles" on page
220.
Load the solvent bottles, waste bottles, and sample vials into the turret.
6850 Automatic Liquid Sampler—See
6850 Automatic Injector—See
Load (or program) the GC sequence.
Run the GC sequence.
shows the critical dimensions for sample vials and microvial inserts
5 minimum
3.5
28.2 ± 0.3
Crimp cap sample vial
All dimensions in millimeters
6850 Series II GC User Information
"Preparing the solvent and waste
"Loading the turret" on page
5 minimum
8.2 ± 0.2
maximum
3.5 ± 0.3
11.7 ± 0.2
Automatic Liquid Sampler and Injector
Preparing sample vials
"The sample turret" on page
34.5 mm
Maximum height of a vial with cap
229.
236.