Download Print this page
UNI-T UT201+ User Manual

UNI-T UT201+ User Manual

Ac clamp meter

Advertisement

Quick Links

P/N:110401108829X
UT201+/UT202+/UT202A+
AC Clamp Meter User Manual
Preface
Thank you for purchasing the new AC clamp meter. In order to use
this product safely and correctly , please read this manual thoroughly ,
especially the Safety Instructions part.
After reading this manual, it is recommended to keep the manual at
an easily accessible place, preferably close to the device, for future
reference.
Limited Warranty and Liability
Uni-Trend guarantees that the product is free from any defect in material
and workmanship within one year from the purchase date. This warranty
does not apply to damages caused by accident, negligence, misuse,
modification, contamination or improper handling. The dealer shall
not be entitled to give any other warranty on behalf of Uni-T rend. If
you need warranty service within the warranty period, please contact
your seller directly.
Uni-Trend will not be responsible for any special, indirect, incidental
or subsequent damage or loss caused by using this device. As some
countries or regions do not allow limitations on implied warranties and
incidental or subsequent damages, the above limitation of liability may
not apply to you.
I. Overview
The UT201+, UT202+, and UT202A+ are true RMS AC clamp meters.
They are designed according to EN61010-1 and CAT II 600V/
CAT III 300V safety standards. These meters come with full-featured
protection which ensures users a safe and reliable measurement
experience. Aside from all the normal features of AC clamp meter,
these meters also include high voltage frequency measurement, fast
capacitance measurement, audio visual NCV detection, and plenty
of additional safety features. UT202A+ also comes with live/neutral
wire detection.
II. Features
• True RMS measurement
• Audio visual NCV detection
• Maximum measurable voltage: 600V;
• High voltage frequency range: 10Hz~10kHz
• Current (UT201+/UT202+: 400A, UT202A+: 600A) frequency
response: 50Hz~100Hz; current frequency measurement function
• Large capacitance (4mF) and temperature measurement (UT202+ only)
• Very large capacitance (60mF), low voltage frequency (10MHz),
and live/neutral wire measurement functions (UT202A+ only)
• Large LCD and fast refresh rate (3 times/s)
• Response time for capacitance measurement: less than 3s for 1mF;
about 6s for ≤10mF; about 8s for
60mF
• Full-featured false detection protection for up to 600V (30kVA) energy
surge; overvoltage and overcurrent alarm functions
• The power consumption of the meter is about 1.8 mA. The circuit
has an automatic power saving function. The consumption in sleep
state is <11uA, which effectively extends the battery life to 400 hours.
Warning: Before using the meter, please read the Safety
Instructions
carefully.
III. Accessories
Open the package box and take out the meter. Please double check
whether the following items are missing or damaged.
a) User manual ---------------------------------------------- 1 pc
b) Test leads
-------------------------------------------------
1 pair
c) K-type temperature probe (UT202+ only)
---------
1 pc
d) Cloth bag
--------------------------------------------------
1 pc
If any of the above is missing or damaged, please contact your supplier
immediately.
IV. Safety Instructions
The meter is designed according to EN61010-1, 61010-2-032/033
and electromagnetic radiation protection EN61326-1 safety standards,
and conforms to CAT II 600V, CAT III 300V, double insulation and
pollution grade II.
Note: In case the meter is not used in accordance with the
operating instructions, the protection provided by the meter may be
weakened or lost.
1.Before use, please check if there is any item which is damaged or
behaving abnormally. If any abnormal item (such as bare test lead,
damaged meter casing, broken LCD, etc.) is found, or if the meter
is considered to be malfunctioning, please do not use the meter.
2.Do not use the meter if the rear cover or the battery cover is not
covered up, or it will pose a shock hazard!
3.When using the meter, keep fingers behind the finger guards of the
test leads, and do not touch exposed wires, connectors, unused
inputs, or circuits being measured to prevent electric shock.
4.The function switch should be placed in the correct position before
measurement. It is forbidden to change the position during
measurement to avoid damage to the meter!
5.Do not apply voltage over 600V between any meter terminal and
earth ground to prevent electric shock or damage to the meter.
6. Be cautious when the measured voltage is higher than 60V (DC)
or 30Vrms (AC) to avoid electric shock!
7.Never input voltage or current which exceeds the specified limit. If
the range of the measured value is unknown, the maximum range
should be selected.
8.Before measuring the resistance, diode and continuity online, switch
off the power supply of the circuit, and fully discharge all capacitors
to avoid inaccurate measurement.
9.When the "
" symbol appears on the LCD, please replace the batteries
in time to ensure measurement accuracy. If the meter is not in use for a
long time, please remove the batteries.
10.Do not change the internal circuit of the meter to avoid damage to the
meter and user!
11.Do not use or store the meter in high temperature, high humidity, flammable,
explosive or strong magnetic field environments.
12.Clean the meter casing with a soft cloth and mild detergent. Do not use
abrasives or solvents!
V. Electrical Symbols
Symbol
Description
Caution, possibility of electric shock
Alternating current
Direct current
Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION
Earth (ground) TERMINAL
Warning or Caution
VI. General Specifications
1. Max display: 4099 (UT201+/UT202+), 6099 (UT202A+)
2. Polarity display: Auto
3. Overload display: "OL" or "-OL"
4. Low battery indication: The "
" symbol is displayed.
5. Low battery shutdown prompt: The "Lo.bt" interface appears on the LCD
and lasts for about 10s, the buzzer beeps three times, and the meter
automatically shuts down.
6. Test position error: If the source under test is not placed at the center of
the clamp jaws when measuring current, ±1.0% additional error in reading
will be produced.
7. Drop protection: 1m
8. The maximum size of jaw opening: 28mm in diameter
9. Battery: AAA battery 1.5V×2
10.Auto power off: If there is no operation of the function switch or any
button for 15 minutes, the meter will automatically power off. This function
can be turned off as needed.
11.Dimensions: 215mm×63.5mm×36mm
12.Weight: About 248g (including batteries)
13.Altitude: 2000m
14.Operating temperature and humidity: 0°C~30°C ( 80%RH), 30°C~40°C
( 75%RH), 40°C~50°C ( 45%RH)
15.Storage temperature and humidity: -20°C~60°C ( 80%RH)
16.Electromagnetic compatibility:
RF=1V/m, overall accuracy = specified accuracy + 5% of range
RF>1V/m, no specified calculation
VII. External Structure (Picture 1)
1. NCV sensing end
2. Clamp jaws
3. Hand guard
4. LED indicator
5. Jaw opening trigger
6. Function switch
7. LCD display
8. Function buttons
9. Positive (+) input jack
10.COM (negative -) input jack
VIII. Button Description
1.SELECT Button
In the composite function position, press this button to switch between the
corresponding measurement functions; in the AC/DC/Hz position (UT202+/
UT202A+), short press this button to switch between the AC and DC
functions, and long press (about 2s) this button to enter/exit the Hz
measurement function.
In the NCV/LIVE position (UT202A+ only), short press this button to switch
between the EFHI and EFLo ranges, and long press (about 2s) this button
to enter/exit the LIVE measurement function.
2.HOLD/BACKLIGHT Button
Short press this button to enter/exit the data hold mode, and long press
(about 2s) this button to turn on/off the backlight (the backlight will
automatically turn off after 60s).
3.MAX/MIN Button
Short press this button to enter the maximum/minimum measurement mode
and long press this button to exit (only valid for AC/DC voltage, AC current,
resistance and temperature measurement).
4.REL Button (UT202+/UT202A+)
In the capacitance and voltage positions, press this button to store the
current reading as a reference for future readings. When the LCD display
value is reset to zero, the stored reading will be subtracted from the future
readings. Press this button again to exit the relative value mode.
IX. Operating Instructions
1.AC Current/Current Frequency Measurement (Picture 2)
1) Select the AC current range (4A/6A, 40A/60A or 400A/600A)
2) Press the trigger to open the clamp jaws, and fully enclose one conductor.
3) Only one conductor can be measured at a time, otherwise the
measurement reading will be wrong.
Note:
• The current measurement must be taken within 0°C~40°C. Do not
suddenly release the trigger, as the impact will change the reading
for a short time.
• To ensure measurement accuracy , center the conductor in the jaws.
Otherwise, ±1.0% additional error in reading will be produced.
• When the measured current is ≥400A (UT201+/UT202+)/ 600A
(UT202A+), the meter will automatically sound an alarm and the
high voltage alarm prompt "
• If the LCD displays "OL", it indicates that the current is over range
and there is a danger of damage to the meter.
2.AC/DC Voltage and Voltage Frequency Measurement (Picture 3)
1) Insert the red test lead into the "
the " COM " jack.
2) Turn the function switch to the AC/DC voltage position, and connect
the test leads with the measured load or power supply in parallel.
Note:
• Do not input voltage above 600V. Although it is possible to measure
higher voltage, it may damage the meter.
• Be cautious to avoid electric shock when measuring high voltage.
• When the measured voltage is ≥30V (AC) or ≥60V (DC), the LCD
will display the high voltage alarm prompt "
3.Resistance Measurement (Picture 4)
1) Insert the red test lead into the "
the "
COM
" jack.
2) Turn the function switch to the "Ω" position, press the SELECT
button to select resistance measurement, and connect the test
leads with both ends of the measured resistance in parallel.
1
2
3
4
5
6
Note:
7
• If the measured resistor is open or the resistance exceeds the
8
maximum range, the LCD will display "OL".
• Before measuring the resistance online, switch of f the power supply
9
10
of the circuit, and fully discharge all capacitors to avoid inaccurate
Picture 1
measurement.
• If the resistance is not less than 0.5Ω when the test leads are
short-circuited, please check the test leads for looseness or other
abnormalities.
• Do not input voltage higher than 30V to avoid personal injury.
4.Continuity Test (Picture 4)
1) Insert the red test lead into the "
the "
" jack.
2) Turn the function switch to the "
button to select continuity measurement, and connect the test leads
with both ends of the measured load in parallel.
3) Measured resistance <10Ω: The circuit is in good conduction status;
the buzzer beeps continuously. Measured resistance >31Ω: The
buzzer makes no sound.
Note:
• Before measuring the continuity online, switch off the power supply
of the circuit, and fully discharge all capacitors.
• Do not input voltage higher than 30V to avoid personal injury.
5. Diode Test (Picture 4)
1) Insert the red test lead into the "
the " COM " jack. The polarity of the red test lead is "+" and that
of
the black test lead is "-".
2) Turn the function switch to the "
button to select diode measurement, and connect the test leads
with the positive and negative poles of the measured diode.
3) 0.08V ≤ reading <1.2V: The buzzer makes one beep indicating
the normality of the diode. Reading <0.08V: The buzzer beeps
continuously indicating the damage of the diode. For the silicon
PN junction, the normal value is generally about 500~800 mV.
Note:
• If the diode is open or its polarity is reversed, the LCD will display
"OL".
• Before measuring the diode online, switch off the power supply of
the circuit, and fully discharge all capacitors.
• Do not input voltage higher than 30V to avoid personal injury.
AC current
Current frequency
measurement
measurement
Picture 2
" will automatically flash.
" jack, black into
Hz
Red
Black
Picture 3
" jack, black into
Capacitance
Continuity
Diode
Resistance
Red
Black
Picture 4
" jack, black into
" position, press the SELECT
" jack, black into
" position, press the SELECT

Advertisement

loading

Summary of Contents for UNI-T UT201+

  • Page 1 2.AC/DC Voltage and Voltage Frequency Measurement (Picture 3) Equipment protected throughout by DOUBLE not be entitled to give any other warranty on behalf of Uni-T rend. If 1) Insert the red test lead into the “ ” jack, black into...
  • Page 2 6.Capacitance Measurement (UT202+/UT202A+, Picture 4) 7. Diode 1.AC Current 1) Insert the red test lead into the “ ” jack, black into the “ COM ” jack. Range Overload Protection Range Overload 2) Turn the function switch to the “ ”...

This manual is also suitable for:

Ut202+Ut202a+