Connecting the device

Opening the device
1. Unscrew the cover of the electrical cable compartment. An identification text "FIELD TERMINAL" is provided at the side
of the housing.
Connecting the device
1. Insert the connecting cable through the cable gland
2. Connect the device to the system with the protective conductor connection
3. Connect the wires to the connecting terminals
Observe the polarity when doing this. If necessary, ground the device using the "-" connection by connecting the
"-" connection to the ground terminal
4. If necessary, ground the shield to the screw of the ground terminal
protective conductor connection.
Power supply isolator with installed load
resistor
Auxiliary power
Ground terminal
Figure 4-1
Schematic electrical connection
SITRANS P500 with HART
A5E02307340-03, 04/2013
.
"+" and "-".
.
Connecting terminals
Protective conductor connector/equipotential bonding terminal
Cable entry for auxiliary power/analog output
.
. This is electrically connected with the external
17