Optical Sensor Connection - Barco DICOM THEATER System Manual

Table of Contents

Advertisement

8 For all the devices in the configuration, define (if necessary) and
select a Preset.
9 After selecting the Preset, MediCal Pro starts consistency
calibration automatically.
Please refer to the MediCal Pro User Guide for more information.

OPTICAL SENSOR CONNECTION

The optical sensor can be attached to the screen by means of Velcro,
attached to the rear of the sensor.
The sensor must be connected to the PC through USB.
Note: When you plug in the optical sensor to the PC USB bus for the
first time, you will be asked to install a driver. You can find the driver on
the Dicom Theater application CD-ROM, in the folder Drivers >
Sequel.
To connect the sensor:
1 Before attaching the sensor, be sure to set the sensor patch to the
desired location (see above). You can always change the position
of the patch by means of MediCal Pro.
2 Peel off the protective sheet of the Velcro part that will be attached
to the screen.
3 Attach the sensor to the screen, in the center of the optical sensor
patch.
Landscape projection
System manual DICOM Theater
Portrait projection