Installing Batteries; Specifications - ADC Adtemp ULTRA 417 Instructions For Use

Digital fever thermometer

Advertisement

Insert the tip of the probe no more than 1/2 inch into the rectum. STOP if you meet any
resistance. The sensing unit is on the very tip of the probe and there is no need to insert the
probe deep into the rectum. Dispose of the used probe cover and wash the thermometer as
recommended. The peak temperature should be reached in approximately 2 seconds.
Axillary Use: Wipe armpit with a dry towel. Place the probe
in the patient's armpit (Figure 3) and keep the patient's arm pressed
firmly against the body. The time required for an accurate underarm
temperature may be 2 seconds or longer depending on placement of
the probe and closure of the arm against the body.
Dual Color Display:
• If a high heat flow is detected from the patient's body, the LCD backlight color will turn
red. Otherwise, the LCD backlight color will display a green light.
• The LCD backlight color is an additional alert to the displayed temperature.
WARNING: A high temperature requires immediate medical attention.
Turning the Unit Off: Turn the thermometer off by pressing the multifunction button. If you
forget to turn it off, the thermometer will automatically turn off after 1 minute.

Installing Batteries

1. Switch off the device.
2. Remove the battery cover by sliding it up.
3. Press the latch holding the battery to remove it from the housing.
4. Place new 2032Li 3V coin type battery in the chamber with the positive (+) side up.
5. Slide the battery cover back into place. Make sure the latch is firmly closed.
It is recommended you remove the battery from the thermometer when it will not be used for
a long period of time.
Troubleshooting and Special Indications
The thermometer has been designed to display feedback warnings to ensure proper usage
and accurate temperature measurement. In the event of a malfunction or incorrect temperature
measurement, the following warnings will be displayed:
Display: Flickering LO (unit will beep 7 times)
Indication: The measured temperature is below 95°F (35°C).
1. Turn the unit off, wait 1 minute and then turn on again.
2. Retake temperature using proper technique while ensuring normal room
temperature (between 73°F to 77°F / 23°C to 25°C).
3. If unit still reads LO, the patient has low body temperature.
WARNING! A low body temperature requires immediate medical attention.
Display: Flickering HI (unit will beep 7 times)
Indication: The measured temperature is above 107.6°F (42°C).
1. Turn the unit off, wait 1 minute and then turn on again.
2. Retake temperature using proper technique.
3. If unit still reads HI, the patient has an extremely high temperature.
WARNING! A high temperature requires immediate medical attention.
Display: Er 1 (unit will beep 5 times and power off automatically)
Indication: The thermometer is too cold or too hot to operate.
1. Verify room temperature is within product range 60.8°F to 104°F (16° C to 40° C).
2. Wait 1 minute, retake temperature using proper technique.
Display: Er 2 (unit will beep 5 times and power off automatically)
Indication: Temperature did not register because of poor contact, patient move-
ment or poor placement.
1. Turn the unit off, wait 1 minute, turn it on again and retake temperature
using proper technique.
Display: Er 3 (unit will beep 5 times and power off automatically)
Indication: Unit malfunctioned.
1. Do not use the thermometer – contact Customer Service.
Display: Flickering battery (after displaying the last temperature measurement)
Indication: Low battery.
1. The battery has enough remaining life for approximately 10 more measurements.
It is recommended to replace the battery as soon as possible.
Display: Battery icon is on (unit will beep 5 times and power off automatically)
Indication: Drained battery.
1. The battery is drained. Replace battery.
Cleaning and Disinfecting
• Wipe probe with an alcohol swab or a cotton swab moistened with 70% isopropyl
alcohol before and after each use. After cleaning, wait at least 2 minutes before taking a
measurement for the unit to readjust to room temperature.
• Use only a soft dry cloth to clean the body of the thermometer (if necessary). Never use
an abrasive cleanser or submerge the thermometer in water or other liquids.
• Store at room temperature.

Specifications

Size:
5"L x 1.14"W x .5"H
(12.7cm L x 2.9cm W x .1.3cm H)
Weight:
0.621 oz. (17.6 g) including battery
Display:
LCD, Liquid Crystal Display - 15.9*18.6mm
Battery:
2032Li 3V Coin Battery
Battery Life:
Minimum 1000 full duty cycles
over a 24 month period
Measurement Time:
Oral 2* Seconds / Rectal 2 Seconds
/ Axillary 2 Seconds
Measurement Range: 95°F to 107.6°F (35°C to 42°C)
Accuracy:
98.6°F to102.2°F +/- 0.2°F
(37°C to 39°C +/- 0.1°C)
THE FCC WANTS YOU TO KNOW...
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may
not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause
undesired operation.
LIMITED WARRANTY
ADC warrants its products against defects in materials and workmanship under normal household use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with the date of delivery.
2. Your Adtemp digital thermometer is warranted for two years from date of purchase.
What Is Covered: Replacement of parts, and labor.
What Is Not Covered: Transportation charges to ADC. Damages caused by abuse, misuse, accident, or negligence.
Incidental, special, or consequential damages. Some states do not allow the exclusion or limitation of incidental, special, or
consequential damages, so this limitation may not apply to you.
To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Dr., Hauppauge, NY
11788. Please include your name and address, phone no., proof of purchase, and a brief note explaining the problem.
Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in no case beyond
the original selling price (except where prohibited by law). This warranty gives you specific legal rights and you may have other
rights which vary from state to state.
For Australian Consumers: Our goods come with guarantees that cannot be excluded under the Australian Consumer Law.
You are entitled to a replacement or refund for a major failure and compensation for any other reasonable foreseeable loss or
damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure
does not amount to a major failure.
Grand Bond Electronics Ltd
Room 2208 22/F CCT Telecom Bldg.
11-19 Wo Shing Street
Fotan, New Territories, Shatin
Hong Kong
Medes Limited
EC
REP
5 Beaumont Gate
Shenley Hill, Radlett
Hertfordshire, WD7 7AR
United Kingdom
Read the instruction manual carefully before using
this device, especially the safety instructions, and
keep the instruction manual for future use.
Type BF applied part
IB p/n 93-417-00 rev 2
Measurement Mode: Predictive measurement
Audible Alarm:
Auto Turn Off:
Memory Feature:
Operation
Environment:
Storage
Environment:
Standards:
*2 second response time achieved in water bath test.
55 Commerce Drive, Hauppauge NY 11788
2797
Figure 3
1 beep at start-up. Continuous beep
at measurement completion
60 seconds
Stores last temperature taken
60.8°F to 104°F (16°C to 40°C)
Humidity: 15-95% (noncondensing)
-4°F to 122°F (-20°C to 50°C)
Humidity: 15-95% (noncondensing)
Complies with ASTM E1112-00(2011)
standard requirements
Distributed by: ADC
Product of Israel
Packed in China
Inspected in the U.S.A.
tel: 1-631-273-9600
toll free: 1-800-232-2670
fax: 1-631-273-9659
www.adctoday.com
email: info@adctoday.com
Batteries and electronic devices must be
disposed of in accordance with the locally
applicable regulations, not with domestic
waste.
®