Troubleshooting; Curing Performance Test; Repair Information; Accessories And Spare Parts - 3M Curing Light 2500 M 5560 Instructions For Use Manual

Visible light curing unit
Table of Contents

Advertisement

Available languages

Available languages

Troubleshooting:

Condition
Main indicator light inoperative
Curing bulb inoperative
No time signal tone
Reduced curing performance
Handpiece turns on, immediately hear
5 beeps and handpiece shuts off
Unit overheats. Five consecutive beeps,
bulb shuts off during curing cycle

Curing Performance Test

Lay a clean polyester matrix band on white paper and place a clear or white plastic form on the band.
The form should be a cylindrical hole 3-6 mm diameter through material 2.0-2.5 mm thick. Pack restorati-
ve material into the form, cover with polyester and press flat. Direct the curing light at the top surface
(from less than 2 mm) of the sample and cure for the recommended time. Remove the polyester strips
and examine the condition of cure. The bottom surface should appear glassy and resist scratching by a
sharp explorer, indicating adequate cure. (Consult manufacturer's instructions for other criteria for correct
cure).
If the surface is soft or dull, repeat the test procedure using incrementally longer curing times until ade-
quate cure is attained, and adjust curing time accordingly. If the adjusted time exceeds 200% of normal,
verify that the restorative material has not degenerated, or correct curing light problems as directed in the
Troubleshooting table.
Repair
Obtain qualified repair service only from 3M Dental Products authorized service facilities.
Contact:
3M Health Care Service Center
Suite 200 Bldg. 502
3350 Granada Ave. North
Oakdale, MN 55128
(800) 292-6298
Replacement Parts and Accessories
Item
Bulb/12V, 75W
Light guide - 8 mm diameter
Light guide - 13 mm diameter
Mounting Kit
Eyeshield - 8 mm
Eyeshield - 13mm
Specifications
Note: This manual is used for all versions of the 3M Curing Light 2500, M5560.
Operation of equipment - Continuous Duty
- Unit will operate continuously for 5 minutes before automatic shutoff, at which time
the unit may be restarted immediately
Remedy
1.
Check electrical supply.
2.
Send unit to an authorized service center.
1.
Replace bulb.
2.
Send unit to an authorized service center.
1.
Send unit to an authorized service center.
1.
Check intensity with hand held radiometer.
2.
Clean light guide exit tip. Consult
cleaning instructions.
3.
Clean optical filter in nose cone with a soft cotton swab.
4.
Check intensity output.
5.
Check optical filter in nose cone for chips or cracks. If
chips or cracks are present on optical filter, send unit to
an authorized service center.
6.
Replace light guide.
7.
Replace bulb.
8.
Check intensity output.
9.
Send unit to an authorized service center.
1.
Fan possibly blocked, therefore send unit
to an authorized service center.
1.
Let unit cool down for 10 minutes.
Restart bulb. If, after a few minutes the bulb shuts off
again after 5-beep signal, or if the bulb will not restart,
send the unit to an authorized service center.
- Can be purchased from authorized dealer
3M Stock Number
78-8045-4571-9
78-8060-9551-5
78-8060-9552-3
12-2716-0339-6
78-8055-7567-3
78-8060-9550-7
7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents