Permissible operating
+10 °C to +40 °C, ≤85 % relative
conditions
humidity (non-condensing)
Permissible storage
-20 °C to +50 °C, ≤85 % relative
conditions
humidity, 800-1050 hPa ambient
pressure
Power supply
2 x 1.5V
Battery life
For approx. 200 measurements, de-
pending on levels of blood pressure
and pump pressure
Data transfer
2402 MHz – 2480 MHz frequency
band
Transmission power max. 2.8 dBM
The blood pressure monitor uses Blue-
tooth
Compatible with Bluetooth
smartphones/tablets
List of supported smartphones/tab-
lets
The serial number is located on the device or in the bat-
tery compartment.
Technical information is subject to change without notifi-
cation to allow for updates.
This device is in line with European Standard EN
60601-1-2 (In accordance with CISPR 11, IEC 61000-
4-2, IEC 61000-4-3 and IEC 61000-4-8) and is subject
AAA batteries
®
low energy technology
®
to particular precautions with regard to electromagnetic
compatibility (EMC). Please note that portable and mo-
bile HF communication systems may interfere with this
unit.
This device corresponds to the EU Medical Devices
Directive 93/42/EEC, the German Medical Devices Act
(Medizinproduktgesetz) and the standards EN  1060-1
(non-invasive sphygmomanometers, Part 1: General
requirements), EN  1060-3 (non-invasive sphygmoma-
nometers, Part3: Supplementary requirements for
electro-mechanical blood pressure measuring systems)
and IEC  80601-2-30 (Medical electrical equipment –
Part 2-30: Particular requirements for the basic safety
and essential performance of automated non-invasive
4.0
sphygmomanometers).
The accuracy of this blood pressure monitor has been
carefully checked and developed with regard to a long
useful life.
If using the device for commercial medical purposes,
it must be regularly tested for accuracy by appropriate
means. Precise instructions for checking accuracy may
be requested from the service address.
We hereby confirm that this product complies with the
European RED Directive 2014/53/EU. The CE Declara-
tion of Conformity for this product can be found under:
www.beurer.com/web/we-landingpages/de/cedeclara-
tionofconformity.php.
36