SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <boolean>
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <boolean>
This command is used with the E-Series power sensors and N8480 Series
power sensors (excluding Option CFT) to enable and disable autoranging.
When autoranging is ON, the power meter selects the best measuring
range for the measurement. When autoranging is set to OFF, the power
meter remains in the currently set range.
The [SENSe[1]]|SENSe2:POWer:AC:RANGe command disables
autoranging.
If INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to
IMMediate, the range will track the input power if
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO is ON.
If the power meter is not making measurements then autoranging only
occurs when the power meter is triggered.
Syntax
SENS
1
SENS2
Example
POW:AC:RANG:AUTO 0
Reset Condition
On reset, autoranging is enabled.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO?
The query enters a 1 or 0 into the output buffer indicating the status of
autoranging.
9-48
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
POW
:AC
:RANG
1 is returned when autoranging is enabled.
0 is returned when autoranging is disabled.
:AUTO
Space
This command disables
autoranging.
EPM-P Series Power Meters Programming Guide
0|OFF
1|ON
?