Connecting Analog Input Signals

Table 4-3 summarizes the recommended input configuration for both types of signal sources.
Examples:
AI
Ground-Reference
*
Setting
Differential (DIFF)
Signal Source
Non-Referenced
Signal Source
Single-Ended
(NRSE)
Referenced
Signal Source
Single-Ended (RSE)
Analog Input Ground-Reference Settings
*
Refer to the
DIFF modes and software considerations.
Connecting Ground-Referenced Signal Sources
Refer to the
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 4-3. Analog Input Configuration
Floating Signal Sources
(Not Connected to
Building Ground)
Ungrounded thermocouples
Signal conditioning with isolated
outputs
Battery devices
DAQ Device
AI+
+
+
AI–
AI GND
DAQ Device
AI
+
+
AI SENSE
AI GND
DAQ Device
AI
+
+
AI GND
section for descriptions of the RSE, NRSE, and
section for more information.
M Series User Manual
Ground-Referenced
Signal Sources
Example:
Plug-in instruments with
non-isolated outputs
Signal Source
DAQ Device
AI+
+
+
AI–
AI GND
Signal Source
DAQ Device
AI
+
+
AI SENSE
AI GND
NOT RECOMMENDED
Signal Source
DAQ Device
AI
+
+
V
A
V
B
AI GND
Ground-loop potential (V
– V
) are added
A
B
to measured signal.
© National Instruments | 4-11