Use, Maintenance, And Care Of The Devices Electrostatic Discharge - Agilent Technologies 85056A User's And Service Manual

2.4 mm precision calibration kit
Hide thumbs Also See for 85056A:
Table of Contents

Advertisement

Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning
connectors attached to static-sensitive circuits (like those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when
discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration
components and devices under test (DUT), can also carry an electrostatic charge. To prevent damage to the
test set, components, and devices:
• Always wear a grounded wrist strap having a 1 M resistor in series with it when handling components
and devices or when making connections to the test set.
• Always use a grounded antistatic mat in front of your test equipment.
• Always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the
conductivity of your floor, wear a heel strap.
• always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or
test port. You can, for example, grasp the grounded outer shell of the test port or cable connector
briefly.
• always ground the center conductor of a test cable before making a connection to the analyzer test
port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to
the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Figure 3-1
shows a typical ESD protection setup using a grounded mat and wrist strap. Refer to Chapter 6 ,
"Replaceable Parts," for ordering information on ESD supplies.
Figure 3-1
ESD Protection Setup
3-2
85056A

Advertisement

Table of Contents
loading