Warranty

Country specific information can be found at: www.parker.com/termsandconditions

Standards

EN 61326-1:2006, Class B
EN 61010-1:2010 Ed.3+ C1; C2
Conforms to UL STD 61010-1:2012 Ed.3+ R: 29Apr2016
Certified to CSA STD C22.2 No 61010-1-12:2012 Ed.3+U1;U2
Caution: The maximum recommended pressure for the sensors is 60 psi.
If this is exceeded, problems with leakage and functionality can occur.
Installation & Start Up
Installation of the SciLog
®
DINCon
®
DINPres
®
, and DINTemp
®
Modules must be carried out only by
trained personnel in accordance with the relevant regulations and this operations manual.
Make sure that the technical specifications and input ratings of the SciLog
®
sensors are observed. See
the appropriate Sensor Specifications page.
The protection provided by this equipment may be impaired if the SciLog
®
sensors are used in a
manner inconsistent with this manual or for purposes not specified by the manufacturer.
Maintenance & Cleaning
The SciLog
®
DIN Modules are maintenance free. The single-use SciLog
®
sensors come pre-calibrated
from the factory and require no maintenance.
To remove dust, dirt and stains, the outer surfaces of the DIN Modules may be wiped using a soft, non-
fluffing cloth moistened with water. If required, you may also use a mild detergent or 2-propanol.
The single-use sensors may be sanitized with 0.1 Molar NaOH, or 2-propanol. They may be autoclaved
up to 2 times, or gamma-irradiated with exposure up to 45 kGy.