Item
1
2
PCU Assembly — 16 mm Nominal Diameter
Piston Disassembly
1. Unscrew the PCU assembly from the instrument; use the dowel hole if the cylinder
2. Unscrew piston stop (5).
3. The weight carrier is permanently attached to the piston; hold the weight carrier and
Piston Cleaning
4. Use "non-fluffing", non-abrasive, lint-free tissue or absorbent cloth. Hold the piston
5. To remove all traces of contamination, the piston can be cleaned in a suitable solvent.
info@Fluke-Direct.com
Figure 5-2. PCU Assembly - 10mm - Vacuum
Table 5-2. PCU Assembly - 10 mm - Vacuum - Parts List
Description
Piston
Spec
Cylinder
Spec
is tight.
gently withdraw the piston from the cylinder (3).
by the weight carrier, and rub the tissue back and forth along its length.
O-ring seals (where fitted) are Nitrile rubber, and should not be
immersed in solvents, as they will become damaged. They
should be wiped carefully with a new tissue.
Fluke-Direct
PCU Assembly — 16 mm Nominal Diameter
Part
Item
3
W Caution
Maintenance & Servicing
3
2
1
Description
Part
Bonded Seal
3918392
.com
1.888.610.7664
5
gjn32.eps
5-5