Specifications; Measurement; General; Spo 2 - GE TruSat User Manual

Hide thumbs Also See for TruSat:
Table of Contents

Advertisement

TruSat User's Guide

Specifications

Measurement

General

SpO
2

Pulse rate

PI r pulsatile value

2
Applicability: OxyTip+ Adult/Pediatric and AllFit sensors.
4–2
Specifications are nominal and are subject to change without notice.
Pulse oximetry sensors: OxyTip+ sensors only
Method: red and infrared light absorption
Red LED wavelength range: 650 to 670 nm
Infrared (IR) LED wavelength range: 930 to 950 nm
Average power: ≤ 1 mW
Calibrated for functional oxygen saturation
Calibration range: 70 to 100%
Measurement and display range: 1 to 100%
Display resolution: 1%
First reading, full accuracy: ≤ 10 seconds
Accuracy, A rms (root mean square of paired values; previously represented by
± 1 standard deviation)
70 to 100% ± 2 digits (without motion)
70 to 100% ± 3 digits (during clinical motion)
70 to 100% ± 2 digits (during clinical low perfusion)
Below 70% unspecified
SpO
measurement accuracy is based on deep hypoxia studies using OxyTip+ sensors
2
on volunteer subjects. Arterial blood samples were analyzed simultaneously on multiple
CO-oximeters.
NOTE: Accuracy may vary for some sensors; always check the instructions for the
sensor.
Measurement and display range: 30 to 250 beats per minute (bpm)
Display resolution: 1 bpm
First reading, full accuracy: ≤ 15 seconds
Accuracy
30 to 250 bpm: ± 2 digits or ± 2%, whichever is greater, (without motion)
30 to 250 bpm: ± 5 digits (during motion)
30 to 250 bpm: ± 3 digits (during low perfusion)
Measurement and display range: 0.01 to 9.99
Display resolution: 0.01 PI r
2