Table 1-5
Recommended separation distance
Rated maximum output power
of transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance "d" in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where "P" is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
8
Chapter 1 Getting Started
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
800 MHz to 2.7 GHz
d=2.3√P
0.23
0.73
2.3
7.3
23