Table 1-4
Electromagnetic immunity (continued)
Immunity test
Power frequency (50 Hz)
magnetic field IEC 61000–4–8
Conducted RF induced by RF
fields
IEC 61000–4–6
Radiated RF EM fields and
Proximity fields from RF
wireless communications
equipment
IEC 61000–4–3

Recommended separation distance

The HC271/HC271p monitor is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the HC271/HC271p monitor can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the HC271/HC271p monitor as recommended below,
according to the maximum output power of the communications equipment.
IEC 60601–1–2:2014 Test
levels
30 A/m
3Vrms 150 kHz -80 MHz, 6Vrms
ISM and Amateur Radio Bands
10V/m, 80 MHz - 2.7 GHz
80% AM 1kHz
385 MHz (18Hz Pulse
Modulation)
450 MHz (FM+/- 5KHz deviation
1 kHz sine)
710 MHz (217 Hz PM)
745 MHz (217 Hz PM)
780 MHz (217 Hz PM)
810 MHz (18 Hz PM)
870 MHz (18 Hz PM)
930 MHz (18 Hz PM)
1720 MHz (217 Hz PM)
1845 MHz (217 Hz PM)
1970 MHz (217 Hz PM)
2450 MHz (217 Hz PM)
5240 MHz (217 Hz PM)
5500 MHz (217 Hz PM)
5785 MHz (217 Hz PM)
Compliance level
30 A/m
3Vrms 150 kHz -80 MHz, 6
Vrms
ISM and Amateur Radio Bands
10V/m
80 MHz - 2.7 GHz
27 V/m
28 V/m
9 V/m
9 V/m
9 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
9 V/m
9 V/m
9 V/m
Recommended separation distance
Electromagnetic environment
— Guidance
Power frequency magnetic
fields should be at least
characteristic of a typical
location in a typical commercial
or hospital environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of the HC271/HC271p,
including cables, than the
recommended separation
distance calculated from the
equation appliance to the
frequency of the transmitter.
Recommended separation
distance:
d= 1.2 √P
d= 1.2 √P 80 MHz to 800 MHz
d= 2.3√P 800 MHz to 2.7 GHz
Where P is the maximum power
in watts (W), d is the
recommended separation
distance in meters (m), and E is
the immunity test level in V/m.
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the
compliance level in each
frequency range. Interference
may occur in the vicinity of
equipment marked with the
following symbol:
7