Agilent Technologies G6012A User Information

Quiet cover ds

Advertisement

Quick Links

Agilent G6012A
Quiet Cover DS
User Information
Agilent Technologies

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies G6012A

  • Page 1 Agilent G6012A Quiet Cover DS User Information Agilent Technologies...
  • Page 2 Further, to the max- agreement and written consent from  calls attention to an operating imum extent permitted by applicable Agilent Technologies, Inc. as governed by procedure, practice, or the like that, if law, Agilent disclaims all warranties, United States and international copyright...
  • Page 3: Table Of Contents

    Contents Introduction General Information Important Safety Warnings Cleaning Recycling the Product Parts Identification Site Preparation Installation Installing the Oil Tray Installing the Rough Pump Inside Quiet Cover DS Installing the Mist Filter Extension Installing the Hose Installing the Lid Installing the Pump Access Panels Connecting the Power Supply Verifying the Installation Operation...
  • Page 4 Quiet Cover DS User Information...
  • Page 5: Introduction

    Parts Identification 11 Site Preparation 13 This manual describes how to install, operate, and troubleshoot the G6012A Quiet Cover DS for GC/MS, LC/MS and ICP/MS rough pumps. The Quiet Cover DS encloses your rough pump to significantly reduce noise, while two internal fans ensure your pump will not overheat.
  • Page 6: General Information

    The Quiet Cover DS is designed and manufactured under a quality system registered to ISO 9001. Information The Agilent Technologies Quiet Cover DS meets the following IEC classifications: Category II, Pollution Degree 2, Class III. Quiet Cover DS has been designed and tested in accordance with recognized safety standards and is designed for use indoors.
  • Page 7 Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. Table 1 lists the safety symbols that may appear on Quiet Cover DS.
  • Page 8: Important Safety Warnings

    Failure to comply with these precautions violates safety standards of design and Quiet Cover DS’s intended use. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. Never run the pump without also running Quiet Cover DS.
  • Page 9 • Make sure that all peripheral devices are also certified. • Make sure that appropriate cables are used to connect the device to peripheral equipment. • Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance. • Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to...
  • Page 10: Cleaning

    Introduction Cleaning To clean the oil tray: Reach under the front of Quiet Cover DS and pull out the tray. Wipe out excess oil with a clean, lint- free cloth. Wash the tray with a mild soap solution. Recycling the Product For recycling, contact your local Agilent sales office.
  • Page 11: Parts Identification

    Introduction Parts Identification Table 2 lists the major parts of the Quiet Cover DS kit. Figure 1 and Figure 2 depict the parts of the Quiet Cover Table 2 Quiet Cover DS parts Part Quantity Part number Quiet Cover DS G6012-64000 •...
  • Page 12 Introduction See Figure 13 on page 28 for a view of the front panel controls. N O T E Mist filter Pump access panel (both sides) Two fans Kickstand Power cord connection Power supply Figure 2 Quiet Cover DS (back view) Quiet Cover DS User Information...
  • Page 13: Site Preparation

    Introduction Site Preparation Prepare the installation site to meet the criteria in Table 3. Table 3 Site prep information Criteria Description Physical dimensions Weight 24.5 kg (54 lbs) Dimensions Kickstand up: 89 × 39 × 44 cm (35 × 15.4 × 17.3 in) (L × W × H) Power supply AC input 100 to 240 VAC ±10%...
  • Page 14 Introduction Quiet Cover DS User Information...
  • Page 15: Installation

    G6012A Quiet Cover DS User Information Installation Installing the Oil Tray 16 Installing the Rough Pump Inside Quiet Cover DS 17 Installing the Mist Filter Extension 21 Installing the Hose 22 Installing the Lid 23 Installing the Pump Access Panels 24...
  • Page 16: Installing The Oil Tray

    Installation Installing the Oil Tray Your Quiet Cover DS is shipped to you with the oil tray packaged separately. You need to install it prior to using Quiet Cover DS. To install it, slide it into the bottom of Quiet Cover DS, with the deep part of the tray forward.
  • Page 17: Installing The Rough Pump Inside Quiet Cover Ds

    Installation Installing the Rough Pump Inside Quiet Cover DS To install the rough pump inside Quiet Cover DS: Remove the two pump access panels by loosening the thumbscrews on the top and sides. Remove the lid by loosening the thumbscrews on each side....
  • Page 18 Installation For Agilent DS 202 and DS 302 pumps, align the back edge of the vibration foam with the back end of the pump base (Figure 4). Center the pump on the vibration foam. Pump base back edge Vibration foam edge Figure 4 Proper installation of the DS 202/302 pumps Quiet Cover DS User Information...
  • Page 19 Installation For Agilent DS 402 and DS 602 pumps, align the front edge of the vibration foam with the front end of the pump base (Figure 5). Be sure that the sides of the pump base align with the sides of the vibration foam. Pump base front edge Vibration foam front edge...
  • Page 20 Installation Remove the grommet (Figure 6) from the back upper edge of the Quiet Cover DS frame and place the grommet around the pump power cord. Grommet Figure 6 Grommet location on Quiet Cover DS Plug the pump power cord into the pump, then slide the grommet with the power cord back into the groove on the back of the frame (Figure 7).
  • Page 21: Installing The Mist Filter Extension

    Installation Installing the Mist Filter Extension The Quiet Cover DS requires that the rough pump use a mist filter extension. If a suitable extension is not already installed, install it as follows: If necessary, shut off the rough pump. Remove the existing mist filter, if installed. Install the mist filter extension, the new seal, and clamps onto the outlet port.
  • Page 22: Installing The Hose

    Installation Installing the Hose Install the pump hose to the rough pump (Figure 9). Figure 9 Pump hose connected to the rough pump (DS 302 shown) Refer also to your rough pump documentation for more information. Quiet Cover DS User Information...
  • Page 23: Installing The Lid

    Installation Installing the Lid To install the lid, place the lid on the base. Make sure the studs (posts) on the inner wall of the top lid align with the grooves along the frame, and the retained thumbscrews align with the screw holes. Tighten the retained thumbscrews on both sides (Figure 10).
  • Page 24: Installing The Pump Access Panels

    Installation Installing the Pump Access Panels To install the pump access panels, place the access panels on the lid and base as shown in Figure 11. Align the retained thumbscrews with the screw holes along the frame and lid. Tighten the retained thumbscrews on both panels. Pump access panels Retained...
  • Page 25: Connecting The Power Supply

    Installation Connecting the Power Supply Connect the power supply to the back of Quiet Cover DS by twisting the ring on the circular connector of the power cord (Figure 12). Plug in power supply cord here Figure 12 Connecting the power supply Plug the pronged end of the power cable into the wall outlet.
  • Page 26: Verifying The Installation

    Installation Verifying the Installation To verify that you installed Quiet Cover DS correctly, press the on/off switch located on the right side of the front panel. • The green LED should light and the fans should start turning. • The temperature alarm should be off and the red light should not be lit.
  • Page 27: Operation

    G6012A Quiet Cover DS User Information Operation Front Panel Controls and Indicators 28 Handling Temperature Alarms 29 Changing Rough Pump Oil 30 Using the Kickstand 33 To operate Quiet Cover DS, you will need to: • Understand the controls and indicators on the front panel •...
  • Page 28: Front Panel Controls And Indicators

    Operation Front Panel Controls and Indicators Figure 13 shows the location of the On/Off switch and indicators of the Quiet Cover DS front panel. Thumbscrew (to release access door) On/Off switch Power LED (Green) Temperature Alarm LED (Red) - indicates excessive heat when lit Thumbscrew (to release access door) Figure 13 The front panel of Quiet Cover DS Quiet Cover DS User Information...
  • Page 29: Handling Temperature Alarms

    Operation Handling Temperature Alarms If the air flow to the Quiet Cover DS is restricted or if the room temperature is above 35 °C (95 °F), the red LED on the front panel lights and the temperature alarm sounds. When this condition occurs, remove the panels to cool the pump.
  • Page 30: Changing Rough Pump Oil

    Operation Changing Rough Pump Oil Periodically, you must change the rough pump oil. The oil tray of the Quiet Cover DS is designed to collect the oil as it drains from the pump. Please refer to your pump documentation for detailed information regarding changing your pump oil.
  • Page 31 Operation Remove the oil drain plug from your rough pump (Figure 14). Refer to your pump documentation. Sight glass Oil drain plug Figure 14 DS 302 pump through access door (other DS models are similar) Reinstall the drain plug when the oil has completely drained.
  • Page 32 Operation Remove the access panel over the oil fill plug and refill the pump (Figure 15). Refer to the pump documentation for more information. Oil fill plug Figure 15 Oil fill plug location (DS 302 shown) Replace the pump access panel and close the access door. Quiet Cover DS User Information...
  • Page 33: Using The Kickstand

    Operation Using the Kickstand To tilt the Quiet Cover DS: Make sure the Quiet Cover DS is on a level surface. Lock the two front locking casters to secure the Quiet Cover DS (Figure 16). Front locking casters Figure 16 Locking the front casters Quiet Cover DS User Information...
  • Page 34 Operation Release the kickstand from the retaining clip (Figure 17). Retaining clip Kickstand Figure 17 Releasing the kickstand With the kickstand released from the clip, step down on the horizontal bar until the Quiet Cover DS is in its tilted position (Figure 18).
  • Page 35 Operation For larger pumps, lift gently on the lid above the fans while stepping on the kickstand to help tilt the Quiet Cover DS. The kickstand’s horizontal bar must sit flat on the floor when Quiet C A U T I O N Cover DS is tilted.
  • Page 36 Operation Quiet Cover DS User Information...
  • Page 37: Troubleshooting

    G6012A Quiet Cover DS User Information Troubleshooting Table 4 describes solutions to problems that may occur with Quiet Cover DS. If your problem is not listed or the solution listed does not work, consult your pump documentation or contact your Agilent service representative.
  • Page 38 The Service Manual (part number G6012- 90020) contains part replacement information (procedures performed by Agilent service representatives). Visit www.agilent.com/chem to download a copy of the Service Manual and look for any other current information. © Agilent Technologies, Inc. 2012 Printed in China *G6012-90010* G6012-90010...