Setting Of Tty To Active/Passive; Setting Of Ser. 1; For Local Printer Or Chromlan; Adjustment Of Transmission Rate (Baud Rate) - Siemens PGC 302 Edition II Manual

Process gas chromatograph
Table of Contents

Advertisement

Setting of SER. 1 and SER. 2
to TTY or RS232
The interfaces SER. 1 and SER. 2 can be set to TTY or
RS232 using DIP switches on the PCPU-2.
Interface SER. 2
Switch
On/off
DIP 2
OFF
DIP 2
ON
Interface SER. 1
Switch
On/off
DIP 4
OFF
DIP 4
ON

Setting of TTY to active/passive

The interfaces SER. 1 and SER. 2 can be operated as
TTY interfaces. The plug-in jumpers X8 to X11 can be
used to set the current source separately for the transmit
and receive loops. Also refer to the register Networking:
Various serial interfaces.
Interface
SER. 1
Transmitter
SER. 1
Receiver
SER. 2
Transmitter
SER. 2
Receiver
aktive
The delivered status for both interfaces is passive. This
setting also acts on the terminals for interfaces A and B.

Setting of SER. 1

for local printer or ChromLAN

If the driver UNIDRV1 is loaded for the interface
SER. 1, a DIP switch on the PCPU-2 can be used to set
whether a local message printer is connected to this
interface or whether a ChromLAN is to be operated
there.
Switch
On/off
DIP 3
OFF
DIP 3
ON
3VERDRAH.CHP, Stand Dezember 22, 1999
Meaning
TTY
RS232
Meaning
TTY
RS232
Jumpers
X8
X9
X10
X11
passive
1
1
Meaning
Local printer
ChromLAN
Adjustment of transmission rate
(baud rate)
The transmission rates can be adjusted in three different
manners:
Control unit interfaces
The interfaces com1: and com2: of the control unit can
be adjusted using the software installation program.
Refer to The software: Parameterization of control unit.
PCPU-2 interfaces SER. 1, 3, 4
The baud rates can only be adjusted using the configu-
ration software. A matching interface driver must be
programmed.
PCPU-2 interface SER. 2
The baud rate can only be adjusted on the DIP switch
on the PCPU-2. The DIP switches are scanned by the
driver software.
Switch
On/off
DIP 1
OFF
DIP 1
ON
DIP
switch
Station x1
Station x16
X 9
SER. 1
X 8
X11
SER. 2
X10
SER. 3
X6
SER. 4
X7
IMPORTANT:
When delivered, the DIP switch is set to
00000000. Switches 5 to 8 must always be
switched off in normal operation. Only positions
xxxx0000 are permissible when operating the
chromatograph.
ChromLAN networks
Meaning
9600 baud
4800 baud if TTY
38400 baud if RS232
23