5. Remove each tube from the pump assembly.
a. Push the thumb tab on the pump arm to the left as far as possible.
b. While holding the thumb tab in this position, pull the plastic cuff on the tube until
c. Discard the tube in an appropriate waste container.
Figure 6-10. Removing Tubing from Pump Assembly
6. Remove the new tubes from the Hematek Pump Tube Set.
Each tube is numerically coded to correspond with the numbers shown on the Stain
Pak.
7. Attach the coded end of the new tube to its respective cannula.
a. Replace the cannula if it appears damaged or bent using the Hematek Cannula Set
b. If you are re-using any cannulas, make sure the cannula previously used for stain
Section 6: Minor Replacements and Adjustments, Page 6—9
the tube is completely removed from the pump arm.
Tube
stain tube
buffer tube
rinse tube
(PN 4483).
is used only with the stain tubing.
Code
1
2
3