SWR DensFlow Operating Instructions Manual

Flow-measurement for densephase-conveying

Advertisement

Quick Links

BA_DF_0003_E_07_04

Advertisement

Table of Contents
loading

Summary of Contents for SWR DensFlow

  • Page 1 BA_DF_0003_E_07_04...
  • Page 2: Table Of Contents

    MOUNTING AND INSTALLATION Delivery Range Auxiliary Mounting of the Measuring Pipe Overview of the Connection of the Senor Pipe and Evaluation Unit COMMISSIONING MENU STRUCTURE OF DENSFLOW MENU PARAMETERS OF THE SYSTEM IN DETAIL MAINTENANCE WARRANTY TROUBLE SHOOTING TECHNICAL DATA...
  • Page 3: System Overview

    Sensor (Measuring Pipe) Evaluation Unit 2 Function • DensFlow is a measuring system especially developed for measuring the flow rate of conveyed solids in densephase. • DensFlow is working according to the latest microprocessor technology. By special capacitive linking of an electromagnetic wave a homogeneous measuring field is produced in the pipe.
  • Page 4: Safety

    3 Safety • The measuring system DensFlow was designed, built and tested to be safe and was shipped in safe condition. Nevertheless persons or objects may be endangered by components of the system if these are operated in an inexpert manner. Therefore the operational instructions must be read completely and the safety notes must be followed.
  • Page 5: Mounting And Installation

    4 Mounting and Installation Delivery Range • Measuring instrument in a 19"-rack system. • Sensor for installation into the pipe. • Seal-ring for adjustment to the pipe. • Operating instructions. Auxiliary • Appropriate wrench or ring wrench for screwing. • Tools for adjusting the wiring.
  • Page 6 WARNING ! Before the installation you have to check, if there is a burr, a disalignment or a seal in the intersection of the pipe and measuring sensor. If so, these resistances in the pipe must be removed. Fig. 4: Installation of the sensor accommodation. •...
  • Page 7: Overview Of The Connection Of The Senor Pipe And Evaluation Unit

    Overview of the Connection of the Senor Pipe and Evaluation Unit Fig. 5: Wiring of the Sensor Pipe and Measuring Instrument A maximum length of 300 m of the sensor cable should not be exceeded. A 5-wired cable is needed between sensor and evaluation unit.
  • Page 8 Electrical Connection • Fig. 6: Electrical Connection Evaluation Unit Terminal No. Connection Connection of the Supply Voltage 2a/c + 4a/c Input Supply Voltage +24 V DC 6a/c + 8a/c Input Supply Voltage GND Connections RS 485 Data A RS485 RS 485 Data B Output 4..20mA + Throughput Output 4..20mA –...
  • Page 9: Commissioning

    The correct cabling between sensor and the evaluation unit. • The correct adjustment of the sensor pipe. Commissioning DensFlow For start-up the sensor has to be calibrated and parameterized to each product, which will be measured. It is necessary to assign the mass flow to the display and initial value.
  • Page 10 Average The measuring range filter is used for the adjustment to slower working devices or for a continuous output of the analog output. Menu point 2.3 for velocity and Menu point 2.4 for density. Storage Adjusted values are confirmed by pressing the ENTER- button.
  • Page 11: Menu Structure Of Densflow

    6 Menu Structure of DensFlow 1.1 Display Density 1.2 Display Throughput 1.3 Display Velocity 2. Options 1.4 Display Totalisator 2.1 – Density measuring range 2.2 – Throughput measuring range 2.3 – Average velocity 2.4 – Average density 2.5 – Pipe diameter 2.6 –...
  • Page 12: Menu Parameters Of The System In Detail

    7 Menu Parameters of the System in Detail 1. Display of the Measured 2. Options 3. Diagnosis Values 1.1 - Density 2.1 - Density measuring range 3.1 - A/D-1 (Density) [kg/m³] 1.2 - Throughput 2.2 - Throughput measuring range 3.2 - A/D-2 (Velocity) [kg/h] 1.3 - Velocity 2.3 - Average velocity...
  • Page 13 2. Options: OPTIONS Press ENTER Button Entry of the density measuring range in Density Range 50 kg/m³ - steps. [Kg/m³] Entry of the throughput range in 100 kg/m³ - steps. Throughput Range Final value of measuring range [kg/h] = 20mA 20000 Kg/h Entry of the average time for the Average V...
  • Page 14 3. Service menu: Service menu To get into the diagnostic mode press ESC and ENTER button at the same time ADC 1 (Density) 3.1 Display current value A/D-converter 1 (Density) I=0.0mA 3.2 Display Voltage level A/D-converter 2 (Velocity) ADC 2 (Velocity) I=0.0mA 3.3 Selection Current value for test purposes.
  • Page 15 Offset Calibration DAC-Calibration for 4 mA output 2 (density) Of's DAC2 DAC-Calibration Span Calibration V DAC1 for 20 mA output 1 (throughput) DAC-Calibration Span Calibration V DAC2 for 20 mA output 2 (density) 3.8 Slave Adress Slave Adress select slave adress for ModBus - comunication Totalisator TOTALISATOR with the totalisator function it is possible to monitor...
  • Page 16: Maintenance

    In case of a defect during the warranty period, defective components are repaired or are replaced free of charge. Replaced parts turn into the property of SWR. If desired by the customer that the parts should be repaired or replaced in its factory, then the customer has to take over the costs for the SWR-service staff.
  • Page 17: Technical Data

    11 Technical Data Sensor Pipe Housing: Steel St52, powder-coated (stainless steel 1.4541) option NW 10..250, Flange DIN 2576 Inner pipe: Ceramics, POM, PTFE Protection category: IP65 Operating temperature: Sensor pipe: -20..+ 120 °C Option: -20..+ 220 °C Sensor electronic: 0..+ 60 °C Max.