Standpipes - Siemens XRS-5 Compact Operating Instructions

Ultrasonic transducers for sitrans
Hide thumbs Also See for XRS-5:
Table of Contents

Advertisement

Appendix A: Applications
A.1 Liquids applications
A.1.3

Standpipes

Open Channel Flow Measurement
In many applications, access must be made via a standpipe. In such cases, a Siemens
flange adapter can be used to mate to the flanged standpipe. Another option is to hang the
transducer from a blind flange, respecting chemical compatibility.
The standpipe length should be as short and the diameter as large as possible. As a rule of
thumb, the -3 dB cone of the sound beam should not intersect the standpipe wall in
applications opening into a vessel or larger area. Otherwise, additional blanking will be
required to compensate for the interference zone created by the opening.
Note
When using a stilling well, make sure there is no build-up, welds, couplings, or other debris
on the inside of the well wall. This can affect reliability of level measurement.
No additional blanking required
42
Flanged transducer
No intersection
Vessel
No vessel
No additional blanking required
Compact Operating Instructions, 01/2018, A5E43390688_AA