2.10 Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available
method.
Procedure
Read the value for Mass Flow Rate on the transmitter display.
Menu > Operations > Process Variable Values
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
Connect to the transmitter with the Field Communicator and read the value for Mass Flow Rate.
Online > Overview > Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.
Related information
Flow measurement problems

2.11 Verify the zero

Display
Menu > Service Tools > Verification & Calibration > Meter Zero > Zero Verification
ProLink III
Device Tools > Calibration > Smart Zero Verification and Calibration > Verify Zero
Field Communicator
Service Tools > Maintenance > Calibration > Zero Calibration > Perform Zero Verify
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field
zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the
following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.