Installation; Installation Of Transmitter; Installation Of Sensor - Siemens Sitrans F M Magflo Operating Manual

Electromagnetic flowmeters
Hide thumbs Also See for Sitrans F M Magflo:
Table of Contents

Advertisement

® ® ® ® ®
SITRANS F M MAGFLO
2.

Installation

2.1 Installation of
transmitter

2.2 Installation of sensor

Hazardous area
4
2.
Installation
SFIDK.PS.026.E5.02
Installation of transmitter can be remote or
compact mounted.
Category 2 equipment
In addition to the sensor, the MAG 6000 I as
remote or compact mounted may be installed in
Ex zone 1 and zone 2.
The sensor must always be completely full with
liquid.
Therefore avoid:
• Installation at the highest point in the pipe
system
• Installation in vertical pipes with free outlet
For partially filled pipes or pipes with down-
ward flow and free outlet the flowmeter should
be located in a U-tube.
Installation in vertical pipes
Recommended flow direction: upwards. This
minimizes the effect on the measurement of
any gas/air bubbles in the liquid.